Joerns® Camtec Series Bed Frames

RC 1000 Bed

© 2016 Joerns Healthcare • 6110284 RevB • 16-2016

Caution:

 Drainage Bag Loops are not intended for use 

as tie downs.

Warning: Possible Injury Or Death.

 Bed safe 

working load is 1000 pounds. This is total weight 

counting resident/patient, mattress, bedding and any 

other equipment or persons likely to be on bed. Do 

not exceed 1000-pound safe working load. Exceeding 

the safe working load could result in property damage, 

injury or death.

Warning: Possible Injury Or Death.

 Do not use any 

Side Rails/Assist Devices until you verify they are 

locked in place. Failure to lock assist devices may 

result in injury or death.

Notice:

 Make sure the power cord is unplugged from 

the power outlet prior to disconnecting/connecting any 

foot panel cables. Failure to ensure bed is not powered 

when disconnecting/connecting the foot panel cables 

may result in damage to the bed’s electronic controls.

Warning: Possible Injury Or Death.

 If using 

accessories not manufactured, marketed, or provided 

by Joerns Healthcare for Camtec beds, consult with 

the manufacturer for compatibility and limitations prior 

to use. Failure to do so may result in injury or death.

Bed System Entrapment Information

In April 1999, the U.S. Food and Drug Administration 

(FDA) in partnership with representatives from the 

hospital and post-acute bed industry, including 

Joerns Healthcare, national healthcare organizations, 

resident advocacy groups, and other federal agencies 

formed the Hospital Bed Safety Workgroup (HBSW). 

The workgroup’s goal is to improve the safety of 

bed frames for patients and residents in all health 

care settings who are most vulnerable to the risk of 

entrapment. The efforts of the FDA and the HBSW 

culminated in the FDA’s release of recommended 

guidelines intended to reduce the risk of entrapment, 

including dimensional limits for critical gaps and 

spaces between bed system components and clinical 

guidance for assessment and implementation of bed 

side rails in various health care settings.

Entrapment zones involve the relationship of 

bed components often directly assembled by the 

healthcare facility rather than the manufacturer. 

Therefore, compliance is the responsibility of the 

facility.

As the leading manufacturer of long-term care beds 

and a frontrunner in addressing this critical issue, 

Joerns Healthcare can offer you the expertise, 

assistance and products to bring your facility into 

compliance.

Joerns

®

 Compliance Solutions

Matching the right bed components in order to meet 

regulatory guidelines can be complex.

That is why Joerns offers a wide array of compliance 

options. We assist customers in selecting compliant 

accessories recommended for their specific bed 

model.

For More Information

To learn more about compliance options with Joerns 

products, visit our website at www.joerns.com, or 

contact our Customer Care reps at 800-826-0270 and 

ask for free informational publications.

To learn more about entrapment zones, assessment 

methods and guidelines concerning entrapment, 

contact Joerns Healthcare at 800-826-0270 or 

consult the FDA website:  http://www.fda.gov/

MedicalDevices/ProductsandMedicalProcedures/

GeneralHospitalDevicesandSupplies/HospitalBeds/

default.htm. 

Содержание Camtec RC1000

Страница 1: ...User Service Manual Joerns Camtec Series Bed Frames RC 1000 Bed Expandable Split frame Bariatric Bed Model RC1000 To avoid injury read user s manual before using The Caregiver s Choice TM...

Страница 2: ...nt should not extend below mattress support platform Pendant should not be placed in an oxygen enriched environment such as an oxygen tent Use of electrical circuits in an oxygen enriched environment...

Страница 3: ...group HBSW The workgroup s goal is to improve the safety of bed frames for patients and residents in all health care settings who are most vulnerable to the risk of entrapment The efforts of the FDA a...

Страница 4: ...low it to pass through a 40 door opening The frame is constructed of structural steel The frames lowest position is approximately 16 from the floor to the mattress support platform The frame in the hi...

Страница 5: ...30 to 95 Non Condensing Atmospheric Pressure 700 hPa to 1060 hPa Storage and Shipping Conditions Ambient Temperature 30 C to 50 C Relative Humidity 10 to 95 Atmospheric Pressure 500 hPa to 1060 hPa Be...

Страница 6: ...two halves as indicated 8 Transfer the head and foot end to their new location Keep the head and foot section matched together per serial number ID labels Transfer the head board and foot board to the...

Страница 7: ...e when raising or lowering the bed damage to the bed may result 2 Plug the power cord into a grounded 115 volt socket having a capacity of at least 15 amps Make sure light on control module is illumin...

Страница 8: ...t the power before using the crank The pendant should not be used during hand cranking Adjusting the Bed Width Caution Extend all four sections of each side prior to adjusting bed position or side rai...

Страница 9: ...ajor service and repair must only be performed by authorized Joerns Healthcare Technical Support which you may contact at 800 966 6662 Troubleshooting Problem Possible Cause Solution No functions oper...

Страница 10: ...rays C Check plug for bent or loose prongs 12 Check that Pendant cord has no damage and that it is secured properly to the control box 13 Check to ensure that all decals are affixed at proper location...

Страница 11: ...Lubricate the actuator drive screws using a heavy duty grease such as Drydene 5871 or equivalent This should be done approx every 6 months 24 If bed is a Scale Bed check that the scale functions are o...

Страница 12: ...12 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Wiring Diagram...

Страница 13: ...13 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Scale System Assembly...

Страница 14: ...14 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Main Frame Assembly...

Страница 15: ...15 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Mattress Support Assembly...

Страница 16: ...16 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Caster Base Assembly...

Страница 17: ...17 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Label Location...

Страница 18: ...18 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Labels Used on Bed...

Страница 19: ...16 x 1 LG 17 902518 0 Nut Hex 16 Elastic Lock 18 300591 0 Assembly Foot Board Support 19 300592 0 Assembly Head Board Support 20 106167 0 Emergency Crank Clamp Bracket 21 106166 0 Emergency Crank Bra...

Страница 20: ...2 58 903809 0 Cable Hi Lo Foot End Motor Split to Control Box 59 903812 0 Hi Lo Head End Actuator 60 903803 0 Hi Lo Foot End Actuator 61 104592 0 Spacer x 9 16 62 901007 0 Screw 10 32 x Round Head Ph...

Страница 21: ...ease Plunger 98 904342 0 Spring Compression 99 904343 0 Caplug COF 100 900016 0 Nut Hex 20 Elastic Lock 101 900258 S Washer Lock Split SS 102 900991 0 Washer Nylon x x 1 16 103 901071 0 Screw 20 x LG...

Страница 22: ...e ID RC750 RC1000 Left Foot 131 904447 0 Label Slide ID RC750 RC1000 Right Foot 132 904448 0 Label Slide ID RC750 RC1000 Mid Section 133 904449 0 Label Slide ID RC750 RC1000 Upper Knee Gatch 134 90390...

Страница 23: ...000 0 LB The patient may now be placed on the bed Important When the 0 0 button is pressed everything on the weighing surface will be zeroed out on the display Make sure that all linens pillows etc a...

Страница 24: ...weight you must remove the patient from the bed and re zero the system 1 Press the FREEZE RESUME button The display will read FREEZE patient weight 2 Make all necessary adjustments to bed and or patie...

Страница 25: ...and Disinfecting The readout and transducer for the 1228 Scale System are made of aluminum and steel with a painted surface You need only exercise caution when cleaning the display window as this is...

Страница 26: ...zed in writing by Joerns Healthcare will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts are not covered under the stated warranty but are covered for...

Отзывы: