User-Service Manual

Joerns

®

 Bed Frames

UltraCare

®

XT

To avoid injury, read user’s manual before using.

Содержание UltraCare XT

Страница 1: ...User Service Manual Joerns Bed Frames UltraCare XT To avoid injury read user s manual before using ...

Страница 2: ...guration Failure to do so could result in injury or death Warning Possible Injury Or Death Use a mattress that is properly sized to fit mattress support platform which will remain centered on mattress support platform relative to State and Federal guidelines Joerns Healthcare recommends the use of a mattress with minimum dimensions of 35 inches wide and 6 inches deep Length should match mattress s...

Страница 3: ...e achieved via an externally available backup battery Joerns P N N511 Joerns recommends a minimum of one battery for every twenty five beds per facility Caution This device complies with EMC requirements of IEC 60601 1 2 Radio transmitting equipment cellular phones or similar electronic devices used in proximity of bed may affect bed performance Caution Particular precaution must be considered dur...

Страница 4: ...rather than the manufacturer Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance Solutions Matching the right bed components in order to meet regulatory guidelines...

Страница 5: ... Stops Patented Toroidal Construction US Patent D437143 Patented Lifting Mechanism US Patent 6473922 EasyLift Leg Positioner Battery Pack Cable Electrical Specifications Power 120 VAC Frequency 60 Hz Current Rating 4 0 Amps Classification Class I Type B or Class II Type B Mode of Operation 10 Maximum Duty Cycle 2 minutes on 18 minutes off Circuit Protection Auto Reset Current Sensor Environmental ...

Страница 6: ...ckets into bed sockets 5 Go to final assembly below Final Assembly Step for Head Panel and Frame Mounted Foot Panel 1 Insert head and foot panel as shown with mounting hardware facing head of the bed Figure 3 2 Tighten all screws with panels in place 3 Panels should slip in and out of their sockets easily and securely Installation Of Baseboard Bumper Optional Advance Care Recommended Tool Rubber m...

Страница 7: ... from the top into the mattress stop 4 Replace mattress making sure mattress fills width between stops Also make sure mattress will not compress to less than 1 5 depth under user s load especially when user is sitting on or sliding over mattress side stop IV Sockets Joerns UCXT bed has four IV sockets as standard equipment There are two sockets at head end of bed and two at foot end Figure 7 The I...

Страница 8: ... bed or walls The 12 bumper is identified by a silver label with 12 printed on it 6 To return bed to 4 Advanced Positioning repeat steps 2 and 3 then press bed down Button 6 Figure 20 pg 9 Bed Hi Low lock icon will blink once confirming software is now in 4 Advanced Positioning mode Install correct baseboard bumper Operation Advanced Pendant Control Operation 1 Your UltraCare XT bed comes standard...

Страница 9: ...utton to lower knee 5 Push button to raise bed 6 Push button to lower bed 6 5 2 4 3 1 Figure 18 Figure 16 Figure 17 Standard Staff Control 1 Push button to raise back 2 Push button to lower back 3 Push button to raise knee 4 Push button to lower knee 5 Push button to raise bed 6 Push button to lower bed 7 Push and hold to change lockout status in combination with another function button 2 1 4 6 5 ...

Страница 10: ...at the head When the UltraCare XT is equipped with dual UltraLocks the foot end functions in the same way as the single UltraLock option while the head end has rollers when the UltraLock at the head end is not engaged and swivel casters when engaged 1 There are three possible locations by which to operate the UltraLock on each UltraLock assembly Figure 22 2 To lock the bed using the UltraLock plac...

Страница 11: ...o serviceable parts Overload protection is achieved via current sensing technology which will automatically reset If any bed function fails to operate check to see if bed is overloaded or constrained from movement To Remove Control Box 1 Plugs and outlets are numbered and color coded so they can be easily reinstalled later See Motor and Control Box information 2 Unplug from wall outlet 3 Unplug al...

Страница 12: ...n as original routing 8 Replace cable ties along with cords in their original position and routing to the control box If this step is not followed damage to cords or electric shock may result Troubleshooting Any loss of electrical function should be investigated with the following steps Event Possible Remedies Bed not responding to pendant control 1 Check lockout status 2 Check connection is fully...

Страница 13: ...ghten any loose hardware and contact Joerns Healthcare to order any appropriate service parts 5 Make sure each inspection includes the underside of the bed frame and mattress support platform 6 Visually inspect the bed frame and accessories for any cracking bending or hole enlargement remove the bed from service immediately and replace the affected parts 7 Check wiring for proper connections and d...

Страница 14: ...anel is small enough to prevent resident from getting his her head or neck caught in these locations Failure to do so could increase risk of suffocation or result in injury or death Warning Possible Injury Extender is designed for use with a foot panel Sitting on or sliding over Extender without a foot panel installed could cause extender to fail resulting in injury or death Drainage Bag Holder Wa...

Страница 15: ...s related to this option and follow all instructions below 2 Install Trapeze Adapter according to supplied instructions 3 Install Trapeze Bar assembly onto Trapeze Adapter according to manufacturer s instructions 4 Adjust bed from its highest through lowest positions to insure proper clearance and operation Some trapeze systems may not be suitable for use with the UltraCare XT bed due to its low o...

Страница 16: ...ems suspended from holder to rest on the floor particularly when bed is in its lowest position Otherwise a safety condition may occur or the function of attached items may be impacted Bed Transport Bar Figure 36 1 Bed Transport Bar is designed to ease moving the UltraCare XT bed particularly when it is lowered on to its casters Note The UltraCare XT bed is not designed for use as resident patient ...

Страница 17: ...g yellow light goes off and battery will be fully charged Note We recommend that users charge up battery immediately upon receipt 4 Backup Battery has useful life of about 3 5 years depending upon usage It is recommended to discharge battery as little as possible and to recharge quite frequently by keeping the Backup Battery connected to Battery Charger when not in use to provide the longest life ...

Страница 18: ...ty but are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is found by Joerns Healthcare to be defective Warranty replacement parts are covered by the terms of this warranty until the product s original stated warranty period expires When requested by Joern...