8. Maintenance Schedule & Daily Check List
All Hoyer products are designed for minimum maintenance, however some safety checks and
procedures are required. A schedule of DAILY tasks are detailed below. Daily checks and a six
monthly service, inspection and test will ensure a lift is kept in optimum safe working condition. A
list of spare parts is available upon request.
The LOAD TEST and certification should only be carried out by qualified personnel or an authorised
service agent / dealer.
DAILY CHECK LIST: Joerns Healthcare strongly recommends the following checks be carried out
on a daily basis and before using lift.
• MAKE sure the lift moves freely on its castors.
• MAKE sure the 6-point cradle is free to rotate and swing
• EXAMINE & CHECK that the 6-point cradle is firmly attached to the boom.
• EXAMINE the sling hooks/clips on the 6-point cradle for excessive wear.
• MAKE sure the legs open and close correctly.
• OPERATE the hand control to confirm the boom raises and lowers satisfactorily.
• CHECK the operation of the emergency stop button.
• EXAMINE slings for fraying or other damage. DO NOT use any sling if damaged or if the
sling shows signs of wear.
• CONFIRM the lift is not giving a low battery alarm when the hand control is operated. If the
alarm sounds, DO NOT use, and place on charge immediately.
• MAKE sure the actuator connection points are secure.
MAINTENANCE, INSPECTION AND TEST
Joerns Healthcare recommend a thorough inspection and test of the Hoyer Calibre lift and lifting acces-
sories, slings, scales etc is carried out on a regular basis. Inspection frequency varies depending on
locality, so you must check with your dealer or local government agency as appropriate regarding
how often an inspection is required. The examination and test should be conducted according to
the recommendations and procedures provided in this manual. Joerns Healthcare recommends
maintenance, inspection and certified testing is carried out by authorised service agent / dealers only.
NOTE:
These recommendations are in compliance with the requirements of 1998 No2307 Health and Safety:
The Lifting Operations and Lifting Equipment Regulations 1998. This is a UK regulation. Outside the
UK please check your local country requirement