 Lifting and Repositioning

C-HLA-2 Lift

Inspection of Hydraulic Pump 

Stud

Tools required: 5mm hex wrench, 9/16" combination 

wrench

1.  Observe orientation of “U” bracket (3) and rotation 

limit bracket (4) prior to removing pump from lift.

2.  Remove hydraulic pump from lifter by removing 

bolts at mast and boom attaching points. Place on 

suitable work surface.

3.  Loosen nut (2) by three turns at base of pump 

using the 9/16" combination wrench. If the stud 

turns while loosening, use the 5mm hex wrench to 

prevent it from turning.

4.  Check tightness of stud using the 5mm hex 

wrench. Firmly tighten the stud by turning in a 

clockwise direction when viewed from the bottom.

5.  Verify all components 1, 2, 3, 4, and 5 are correctly 

assembled (Figure 9).

6.  Tighten nut (2) until it lightly contacts the nylon 

washer (1). Loosen the nut (2) by 1/2 turn so 

that the pump rotates freely in “U” bracket (3). 

Check that the stud is still tight using the 5mm hex 

wrench.

7.  Reassemble the pump to lift (reverse of step 

2). Tighten nuts until contact is made with “U” 

brackets of mast and boom. Then loosen by 1/2 

turn. Pump should be free to rotate. Verify that 

“U” bracket (3) and rotation limit bracket (4) are 

correctly oriented to mast per diagram.

8.  Check for free rotation of pump after assembly.

Caution: 

The pump should rotate freely prior to lifting 

a patient.

Caution: 

Place the pump handle in a comfortable 

position prior to lifting a patient. Do not attempt to 

rotate the pump while lifting a patient.

Figure 9

1.  Washer, Nylon 7/8" O.D

2.  Nut, Hex, Locking

3.  “U” Bracket

4.  Bracket, Rotation limit

5.  Washer, nylon, 2" O.D.

1

5

4

2

3

Содержание C-HLA-2

Страница 1: ...Joerns Healthcare User Service Manual Joerns Lifting and Repostioning Model C HLA 2 redefining patient handling To avoid injury read user s manual before using...

Страница 2: ...shipping errors to your supplier Include all equipment identification numbers and group part numbers if any as described above along with a full description of damaged goods Important Precautions War...

Страница 3: ...der and replace all labels that cannot be easily read Caution Use pump to raise boom into position for lifting patient Caution DO NOT lift up on end of boom when positioning to prepare for lifting pat...

Страница 4: ...ion of Hydraulic Pump Stud 7 Operation 8 Leg Spreader Operation Lifter Maintenance Checklist 10 Troubleshooting 11 Hoyer 4 point Sling Information 12 Warranty 14 Specifications Base Adjustable U Maxim...

Страница 5: ...is when lifting a patient from the floor the locks should be engaged during a floor lift 1 Carefully unpack Hoyer Lifter Check for swivel bar 6 point cradle and adjusting lever handle 2 Insert mast in...

Страница 6: ...d secure tightly Figure 6 d Swivel cradle to check for free movement 7 Attach adjusting lever handle into base Figure 7 8 Wheel locks are provided for parking To lock caster step on cam lever on side...

Страница 7: ...2 3 4 and 5 are correctly assembled Figure 9 6 Tighten nut 2 until it lightly contacts the nylon washer 1 Loosen the nut 2 by 1 2 turn so that the pump rotates freely in U bracket 3 Check that the st...

Страница 8: ...imize stability Figure 11 2 Pump hydraulic handle Figure 12 Do not lock rear casters If needed lifter base will move slightly to adjust for balance To Lower Lifter 1 Open hydraulic pressure release kn...

Страница 9: ...e a faulty pump but is a trait common to hydraulic equipment 4 If trouble develops with the hydraulic pump remove it and call your Joerns Healthcare Supplier to receive an RMA Provide serial number lo...

Страница 10: ...bolts and nuts for wear and tightness Boom Inspect all bolts and nuts for wear and tightness Inspect to ensure the boom is centered between the base legs Cradle Check hanger post for wear or damage R...

Страница 11: ...of oil for a jack when completely empty are 7 3 4 oz for 12 stroke 11 1 2 oz for 17 stroke 3 3 4 oz for 6 stroke 12 1 2 oz for 19 stroke Oil leaking from around piston Replace the jack Oil leaking ar...

Страница 12: ...cturer s sling with the C HLA 2 Lifter Note Do not use Hoyer slings with any other manufacturer s lift Patient Size Guide A B Weight inch cm inch cm lbs kg Large 28 70 19 47 198 400 90 191 Medium 24 6...

Страница 13: ...13 Joerns Lifting and Repositioning C HLA 2 Lift Notes...

Страница 14: ...unless performed by or authorized in writing by Joerns Healthcare will void this warranty This warranty applies in the USA only Check with your supplier to find out if international warranties apply P...

Отзывы: