24
JK802/804/808 User's Guide
[TOUCH PANEL]
The LCR meter can work well without touch panel. But you
cannot
Tip
s
type number without touch panel. When you need to input
numbers, the touch panel will be activated even it is shut
down in system setting.
Procedure of setting touch panel
Step 1
Press [SYST] key to enter<SYSTEM CONFIG> page.
Step 2
Select [TOUCH PANEL] field
Step 3
ENABLE
Enable touch panel
DISABLE
Disable touch panel
CALIBRATE Calibrate touch panel
RESET
Reset touch panel data
Procedure to calibrate touch panel
Tips
You will need a screen pen to calibrate touch pane.
Do not use your finger!
Step 1
Press[SYST]key to enter<SYSTEM CONFIG> page.
Step 2
Select[TOUCH PANEL]soft key
Step 3
Select[CALIBRATE]soft key
Step 4
Use a screen pen to click screen softly to start calibration
Step 5
Tap the middle of the circle with a pen on Left-Up corner.
Then
tap the middle of the circle with a pen on the Right-
Bottom corner to finish the calibration.
Step 6
Click on the screen softly to exit.
[BRIGHTNESS]
Four degrees of brightness 30%,50%,70%,100%
Tip
s
If powered by external power, the brightness is 100%.
If powered by battery, the low brightness can make
the
meter work longer.
Содержание JK802
Страница 1: ...Operating Instruction JK802 804 808 Multi channel Handheld Temperature Meter ...
Страница 26: ...26 OFF APO off JK802 804 808User s Guide Tips Timer will be reset when press any keys or touch screen ...
Страница 28: ...RemoteControl 28 recvBuffer ...
Страница 30: ... define program pack Here ...
Страница 34: ......
Страница 38: ...34 JK802 804 808User s Guide 7 Invalid command ...