13
Temperature rise: no function at present
Time setting: instrument clock
Buzzer alarm: the buzzer opens or closes the choice
Power saving: no function at present
Upper limit alarm value: high temperature alarm value. Every 8 channels is an adjustment base,
and can be switched through thermocouple type.
The lower limit alarm value is low temperature alarm value. Every 8 channels are adjusted base
and can be switched through thermocouple type.
Reference value: no function at present
CJC calibration: adjust the channel deviation value, each 8 channels as an adjustment base.
Type of thermoelectric coupler: Type: choose the probe type of the instrument test, each 8
channels as an adjustment base.
Functional selection:
1. When displaying the channel parameters, press the "MENU" button to switch to the
digital bar.Then, through the upper and lower keys to adjust the number you need, and then
switch to the next digit through the left and right keys.When the settings are completed, press
"MENU" to save the settings and exit; press the exit key, do not save the settings and exit.
。
2. Baud rate
:
1200
,
2400
,
4800
,
9600
,
19200
,
38400 According to your different
needs, you can set it on your own. The default is 9600 when you start the machine.
3. Device address: according to the COM port of different PC, that is to choose to several.
The default is 001.
4. Sampling time: how long is it needed to collect, how long it takes to tune up, the
minimum time is 1 second, the maximum is 9999 seconds.
5. Temperature unit
:℃
/
℉
Switch.
6. Automatic screen switching
:
ON/OFF Switch.
7.Time settings: set the year, month, hour, minute and seconds that you need to display.
8. Buzzer alarm: when the temperature is adjusted to "ON", if the temperature exceeds the
upper limit or lower limit set by you, there is a sound of sound. If it is adjusted to "OFF", there is
no sound hint.
9. In the measurement, you can set an upper limit of temperature and a lower limit of a
temperature. When the measured temperature exceeds the range you set, if the buzzer alarm is
in the "ON" state, there will be a sound warning to you. The "+" on the display represents the
positive temperature "-" representing the negative temperature.
10. Type of thermocouple: it can be arranged from 8 to one, and different thermocouple
types are set for a unit.