JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
28
Operation steps for the swap parameter function
Execute the following operations to swap the primary and the secondary
parameters.
1)
Move the cursor to
FUNC
, the following soft key will be displayed.
◼
SWAP
2)
Press
SWAP
to swap the primary and the secondary parameters.
3)
Press
SWAP
to swap the primary and the secondary parameter, which is to
recover the formal setup.
3.5.2
Limit modes of compare function
Compare functions has two limit setup modes for primary parameters as shown in
figure 3-1.
◼
Tolerance mode
Under tolerance mode, set the deviation value of the nominal one (be set in the
NOM
zone) as the compare limit value. Deviation value has two modes:
percentage deviation and absolute deviation.
◼
Sequential mode
Under sequential mode, the range of the test value is the compare limit value.
The compare limit value should be set in the order from small to large.
Figure 3-1 Tolerance mode and Sequential mode
———————————————————————————————————————
Note:
When setting limit values of tolerance mode, the error range should be set
in the order from small to large. If the error range of BIN1 is the largest one, then