JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
27
3.5
<LIMIT TABLE>
Press
[SET]
and then
LIMIT TABLE
to enter into the
<LIMIT TABLE>
page as
the following figure shown.
Compare function can be set on this page. JK2830 can set 3 bin limits of primary
parameters and one of secondary parameters. The tested result can be divided
into up to 4 bins (BIN 1 to BIN 3 and BIN OUT). If the primary parameter of DUT
is within the limit range from BIN1 to BIN3, but the secondary parameter is out of
the limit range, in this case the DUT will be sorted into aux bin. When JK2830
installs the HANDLER interface and it is used in automatic sorting system, the
compare function will be especially useful. The following limit parameter of
compare function only be set on
<LIMIT TABLE>
page.
◼
Test parameter (
FUNC
)
◼
Limit mode of compare function (
MODE
)
◼
Nominal value (
NOM
)
◼
Auxiliary bin ON/OFF (
AUX
)
◼
Compare function ON/OFF (
COMP
)
◼
Low limit of each bin (
LOW
)
◼
High limit of each bin (
HIGH
)
3.5.1
Swap parameter
The swap parameter function can swap the primary and the secondary parameter
in
SWAP
. For example, when the test parameter is Cp-D, the swap parameter
function can change the test parameter as D-Cp. Then user can set 3 pairs of
compare limits for D, but only 1 pair of compare limit can be set for Cp.