
PDL5 User Manual
PDL5-UM-00001 Rev 001
22
The small amount of reflections that BR0 represent can be polarization sensitive,
and multiple reflections can cause interference effects that can make the
reflections very sensitive to temperature.
NOTE: To ensure accurate backreflection measurements below -65 dB,
perform the BR
0
measurement at least once per shift.
The measured backreflection is also affected by losses that can occur between
the meter and the DUT. To compensate for these losses, measure the total
amount of loss between the output of the measurement hybrid jumper and the
input of the DUT (see “LOSS AND POWER MEASUREMENT” section on page 22),
double this value and add it to the backreflection value displayed on the meter.
In the following example, the total loss between the meter and the DUT is 2.0
dB, and the displayed backreflection is -29 dB:
BR = -29 dB + (2 x 2.0 dB) = -25 dB
Absolute accuracy specifications are dependent upon the accuracy of the PDL5
meter calibration, so a calibration verification needs to be performed
periodically. Refer to the “CALIBRATION” section on page 32 for more
information.
Polarization Dependent Loss (PDL), Loss and Power
Measurements
Set-Up for PDL, Loss and Power Measurements
To prepare the meter for PDL, Loss and Power measurements:
1.
Follow the “POWERING UP THE METER” sequence on page 17.
2.
Press the
or
key to select the required output port.
3.
Clean the output port on the front of the meter and the FC/APC
connector of the measurement jumper (see the “CLEANING THE
CONNECTOR ENDS” section on page 35).
4.
Connect the FC/APC end of the jumper to the output port on the
meter.
NOTE: Ensure that you are using the measurement jumper and not the
calibration jumper, which is labeled as such.
5.
Attach the appropriate detector adapter to the detector on the
front of the meter.
6.
Connect the output connector of the measurement jumper to the
detector adapter. This end of the measurement jumper is user-
selected and must be compatible with the input connector of the
DUT.