ESAO
®
Method
CTM Conductimeter
Adaptator ESAO
®
4
Ref :
452 146
ENGLISH
8
The rated capacity number is 2.
The rated capacity 1 varies from 0 to 2 mS/cm
The rated capacity 2 varies from 0 to 20 mS/cm
Click on the Save button to store the new adaptor.
2.7.2 With Generis 5
The adaptor is automatically recognised.
3 TECHNICAL
SPECIFICATIONS
•
Probe input:
by two safety sockets
∅
4mm
•
Rated capacities:
2 mS.cm
-1
, 20 mS.cm
-1
, controlled
by the software
•
Calibration:
in one point by the potentiometer
•
Temperature
compensation:
manual, by a potentiometer from 0 to
50°C
optimum range from 15 to 35 °C
•
Accuracy (excluding the probe):1% of the entire scale
4 After-Sales Service
This material is under a two year warranty and should be returned to our
stores in the event of any defects.
For any repairs, adjustments or spare parts, please contact:
JEULIN - TECHNICAL SUPPORT
Rue Jacques Monod
BP 1900
27 019 EVREUX CEDEX FRANCE
+33 (0)2 32 29 40 50