Installation, Operation, & Maintenance Instructions
IOM J500.35
2021.04
JMT
● +1.440.572.1500
11
Once an acceptable value has been entered, press
Enter
until the cursor is on the last available place. Press
Enter
again and the next property in that menu subsystem will be displayed. If too large or too small a value
is entered for a parameter, it will be rejected with the maximum or minimum value in violation displayed.
Menu System Navigation
The Menu is accessed by pressing the
Enter
button. Details of each submenu are given in subsequent
sections.
Once you have entered a submenu it is necessary to progress through the entire submenu before exiting it.
It
is not possible to use the Back button to exit a Menu.
If you leave the menu system open for 2 minutes with no inputs, it will automatically exit to the
Standard
Operating Screen
.
WARNING
While in the Menu System HART commands will be disabled.
4.2.2 Menu (First Screen)
Device Info
User Installation
Rerange
Flip Transmitter
One Point Calibration
Calibration
Alarms
4.2.3 Menu (Second Screen)
Filters
Reset Factory Configuration
Factory Configuration (Factory Password Protected)
Diagnostics
Exit Menu