2 SPECIFICATIONS AND INSTALLATION REQUIREMENTS
2.4 STANDARD
INSTALLATION
DIAGRAM
Installation Room
Floor space:
2,500 mm (W)
×
3,000 (D) or larger
Ceiling:
2,000 mm or higher
Entrance:
800 mm (W)
×
1,800 mm (H) or larger
When an optional evacuation system is installed, it is necessary to provide a space
of 500 mm or more between the rear of the instrument and the wall of the room.
Typical Installation Layout
710
570
3000
Operation unit
Air
compressor
Computer
710
2500
Main unit
Minimum 500
Minimum 800
Unit: mm
0
500
1000
2-6
TMPM5200-2