6. Restore the scan speed

 

to the

 FAST

 to save the image data on the MO disk. 

 

      Notice: If the scan speed is not restored to the 

FAST, 

the data cannot be saved.

 

 

7. Press the 

INS

 key, and type the file name in the space following the 

SAVE|A:

.       

Press the 

RETURN

 key to save the image data on the MO disk.   

*To confirm the detail of the image capturing, refer to the manufactures instructions.   
 

 

The axis adjustment 

  The adjustment of the movable diaphragm of the objective lens, and the processing. 

 

1. Set the 

magnification to 10,000 times

, and move the marker object on the center of the CRT   

           

screen. 

 

2. Focus the SEM on the marker object, and press the 

WOBBLER

 button. 

 

3. To minimize the shaking of the marker object on the CRT, adjust the movable diaphragm 

using the fine adjustment knob (X, Y). 

  The correction of the astigmatism 

1. Minimize the blur in the image, using the 

FOCUS

 knob.   

 

2. Minimize the blur in the image, using the 

STIGMATOR(X)

 knob.   

3. Minimize the blur in the image, using the 

STIGMATOR(Y)

 knob.   

If the blurry in the image (the astigmatism) is not corrected, repeat the process mentioned 

above (1-3). 

 

  The adjustment of the beam axis 

 

1. Press the 

EOS

MODE

 button once to select the 

SEM1

 mode. 

          Set the magnification to 1,000,000 times using the 

MAGNIFICATION

 knob.   

2. While the ( ) value of the

 CL COARSE

 (

EOS

 menu, 

EOS-1

OPTICS

) is increased

 from 

3

×

10

-12

 to 6

×

10

-10

 using the 

PROBE CURRENT

 knob, maximize the image brightness, 

using the 

GUN ALIGNMENT (X, Y)

 knobs. 

3. Set the magnification to 6,000 times, and repeat the operations mentioned above (2).   

 

4. While the ( ) value of the 

CL COARSE

 (

EOS

 menu, 

EOS-1

OPTICS

from 3

×

10

-12

 to 

6

×

10

-10

 is increased using the 

PROBE CURRENT

 knob, minimize the moving of the 

central object, using the 

GUN ALIGNMENT

 

(X, Y)

 knobs. (When the current is increased, 

center the moved objects.) 

5. Set the value of the 

CL COARSE

 to 

1

×

10

-11

, using the 

PROBE CURRENT

 knob, and then 

perform the adjustments of the movable diaphragm of the objective lens, mentioned above, 