How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)
1.Load the specimen
Load a grid into the rod holder:
USE ONLY THE TOP POSITION
(blue arrow),
Specimen selection on 1 (The rear one is only a reserve one- Spec sel 2). Use the
white wing key with a feather tail (red arrow) to insert the slit over the cover: grab the
cover with the red dots oriented in the long axis, turn a quarter and gently open it.
Insert the grid specimen up. Close the cover and remove the wing key. (Check that all
covers are closed before inserting the rod in the goniometer. Be careful to not lose
the wing key and put it back in its hole on the standard.
If not already done, rotate the tilt of the insertion chamber (goniometer) to the zero
position (scale to be read in the scale on the left side) by releasing the lock
h
(lift it