3.13 GLP Functions

A variety of GLP functions are available via the GLP Setup menu.

A reminder that calibration is due can be set via the Cal Reminder option. Once this option is set a

valid calibration is required to use the 3520 after the time limit has elapsed. This option is

DISABLED by default.

The interval time is set in hours on this menu.

NOTE:

A calibration should be performed immediately after the setting of this value to

reset the clock. This should be done as soon as set up is complete.

The on screen reminder can be accompanied by an audible alarm. This is set using the audible

alarm setting on this menu.

NOTE:

The use of the Calibration Reminder feature will prevent the user from

performing measurements with the 3520 until a valid calibration is carried out.

The user and batch ID can be used to identify sets of samples and a specific user. This information

is printed when data is output to the printer. The batch number is also stored in the results memory.

3.13.1 Security

To control access to set up options and data manipulation functions a security code can be set

using this menu. When enabled a password is required to make any changes to the set up menus.

When the code is ENABLED a user can measure a sample, log data and calibrate the unit, but

cannot change settings within the set up screens until a valid password is entered.

25

Содержание 3520

Страница 1: ...Instruction Manual 352 050 REV B 01 17 pH mV Temperature Meter Model 3520...

Страница 2: ......

Страница 3: ...to the appropriate servicing authority Merci de lire attentivement ces informations avant d installer ou d utiliser cet appareil 1 L appareil d crit dans ce manuel est con u pour tre utilis uniquemen...

Страница 4: ...irgendeiner Weise besch digt wurden muss das Ger t au er Betrieb genommen und gegen weiteren Gebrauch gesichert werden Die St rung sollte der zust ndigen Serviceeinrichtung unverz glich gemeldet werd...

Страница 5: ...adas que se especifican en este manual 3 Cualquier elemento que no se encuentre entre los definidos en los procedimientos de mantenimiento aqu descritos no podr utilizarse en este instrumento La extra...

Страница 6: ...e Values 3 4 Good Practice Guidelines 3 5 Set Up Parameters 3 6 pH Calibration 3 7 Error Codes 3 8 mV Mode 3 9 Performing Measurements 3 10 Status Page 3 11 Results Storage and Display 3 12 GLP Functi...

Страница 7: ...1012 ohms Temperature Measuring Ranges 10 to 105 C 14 to 221 F Resolution 0 1 C 1 F Accuracy 0 5 C 1 F ATC Range 0 to 100 C 32 to 212 F Manual Temp Compensation 0 to 100 C 32 to 212 F Auto Buffer Sel...

Страница 8: ...if required and the pH electrode to the rear panel Temp and pH sockets The unit should be placed on a clean flat surface which is free from drafts and vibrations The units are designed for operation...

Страница 9: ...lly temperature compensated MAN symbol 5 Endpoint symbol this symbol is displayed when an endpoint has been detected 6 Mode tags Each mode tag is highlighted when selected SETUP MODE pH or mV STATUS o...

Страница 10: ...mV in mV mode 3 Print key used to initiate a print 4 Up Arrow used for adjustment during set up to scroll results and to toggle between mV and pH modes Down Arrow used for adjustment during set up to...

Страница 11: ...llows combination pH or redox electrodes to be used 3 Temp Socket 8 pin mini DIN socket This allows the optional Automatic Temperature Compensation ATC probe to be connected 4 Alarm Output 2 x 4mm soc...

Страница 12: ...our as the pH level varies These can be used as a guide to the pH level but can be limited in accuracy and difficult to interpret correctly in murky or coloured samples For greater accuracy the use of...

Страница 13: ...he solutions For accurate and repeatable results it is essential to follow the manufacturers instructions carefully 3 4 Solution Temperature Values The value of all buffer solutions varies with soluti...

Страница 14: ...is level remains above the sample in use 5 DO NOT touch the sensitive glass pH membrane or reference junction during use Excess droplets of solution may be removed by gently blotting with filter paper...

Страница 15: ...to scroll between 0 001pH 0 01pH and 0 1pH mV Resolution allows the user to select the preferred resolution Use the Up Down arrows on the keypad to scroll between 0 1mV and 1mV Language enables the se...

Страница 16: ...rrow keys allow forward and backward movement along the row of digits Press the STO key to accept the value Manual Cal 2 allows Cal 2 manual buffer value to be set using the Up Down keys The Left Righ...

Страница 17: ...to a 4 digit code can be set using the Up Down keys The Left Right arrow keys allow forward and backward movement along the row of digits Press the STO key to accept the value 3 6 4 Alarms Setup Exit...

Страница 18: ...eys allow forward and backward movement along the row of digits Press the STO key to accept the value 3 6 6 Security Setup Select the Security Setup sub menu buy highlighting the option and pressing t...

Страница 19: ...led or disabled by using the Up Down keys which toggle between the two settings Clr Key toggles between Deletes Results Before Deletes All Results and Deletes Results Since Selection is made via the U...

Страница 20: ...and return to the main instrument display Display toggles between Time and Date Selection of the preferred option can be made using the Up Down keys which toggle between the two settings Date Format t...

Страница 21: ...display will show the manually set temperature reading in C or F When manual temperature compensation is being used the annunciator will indicate MAN 2 Immerse the electrode s in the first buffer solu...

Страница 22: ...slope efficiency and then return to the main measuring screen The current calibration information including the Eo value can be found on the Status screens Rinse the electrode s in deionised water 3 C...

Страница 23: ...play will show where 1 3 is the number of calibration points selected in the Calibration Setup menu Value should show the STC value of the buffer used Press the CAL or STO key The display will then sh...

Страница 24: ...the Eo value and the slope efficiency and then return to the main measuring screen The current calibration information including the Eo value can be found on the Status screens Rinse the electrode s...

Страница 25: ...nits from the set value Rinse the electrode s in deionised water Once a successful calibration has been completed the instrument will return to the measuring mode The instrument is then ready to under...

Страница 26: ...icating that corrective action should be taken Recording these values as part of a routine Quality Control program can give a good indication of the condition of the electrode Relative Millivolts This...

Страница 27: ...perature compensation is being used the preferred measurement range should be selected via the Set Up menu b Immerse the electrode into the solution and set the to the solution temperature via the Set...

Страница 28: ...g after a reset or a failed calibration the warning screen will be shown The status page will show the calibration data in the order it was carried out Date time temperature mV and pH readings will be...

Страница 29: ...et data logs when the alarm set point is reached Alarm Clear data logs when the alarm is cleared Alarm Set Clear data logs at the alarm set point and when the alarm condition clears Disabled no data l...

Страница 30: ...select results The most recently stored results will be displayed on the screen If more results are stored than can fit on a single screen then a series of data screens are available The current scre...

Страница 31: ...ing the audible alarm setting on this menu NOTE The use of the Calibration Reminder feature will prevent the user from performing measurements with the 3520 until a valid calibration is carried out Th...

Страница 32: ...in the solution until clean Rinse and repeat as necessary Electrodes which have been allowed to dry out should be soaked overnight in warm distilled water Table of Cleaning Agents for Glass Electrode...

Страница 33: ...essories 5 1 Optional Accessories The following list of items are available as optional accessories for use with the Model 3520 060 406 Dust cover 037 701 IrDA printer supplied with roll of thermal pa...

Страница 34: ...on the rear panel 9 way D type connector The connections are as follows DCD 1 LINKED TO DTR AND DSR RXD 2 INPUT TO 3520 TXD 3 OUTPUT FROM 3520 DTR 4 LINKED TO DCD AND DSR GND 5 DSR 6 LINKED TO DCD AND...

Страница 35: ...trol and software XON XOFF flow con trol Pressing the PRINT key outputs from the RS232 interface Sending an ASCII D to the 3520 causes a printout of the current displayed reading plus sample number Se...

Страница 36: ...ate of the stored readings will be displayed An asterisk indicates that manual temperature compensation is being used A reading in the Relative mV mode will be indicated by an R Each reading will be i...

Страница 37: ...Replace electrode Unable to recognise Using correct buffer set Is the buffer type correct Use AUTO for buffers Jenway supplied buffers Replace buffers Use BNC cap to test 3520 see 7 2 Replace electrod...

Страница 38: ...isplay will momentarily show E2 RST 4 If this does not resolve the problem please contact the manufacturer or your local distributor 7 2 Functional check The measurement function of the meter can be c...

Страница 39: ...declared IEC EN 61010 1 2001 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements Electrical equipment for measu rement control and labor...

Страница 40: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www jenway com...

Отзывы: