innoCon 6800 Monitor Manual
Page 15
Manual Calibration
1.
Enter calibration menu, and select “Manual Calibration”, and press “Enter”
Wash and dry The electrode and put it into the standard solution, enter the
standard solution value, after the reading value is stable, press the "OK" key to
complete the calibration.
Restore
1. Enter calibration menu, and select ”Restore“ and press ”Enter”
Manual Calibration
pH Value :
Current measurement pH value
Input current temperature of standard solution ,
optional range: 0~60.0℃
Make sure input the value of standard solution by manual.
Current measurement pH value
Input current temperature of standard solution ,
optional range: 0~60.0℃
Make sure input the value of standard solution by manual.
This function will restore the calibration value to the factory state
Input Temp.:
Input Buffer1:
Select Buffer and Press ENTER
6.97pH
25.0℃
7.00pH
1
▲
▲
1
Clean and dry the electrode and put it into the first point of the standard solution, enter the
current standard solution temperature (at MTC only), select the first point the standard solution,
press Enter to start cabration automatically, when stable, go to next page automatically
Manual Calibration
pH Value :
Input Temp.:
Input Buffer2:
Select Buffer and Press ENTER
4.23pH
25.0℃
4.01pH
1
▲
▲
Auto Calibration
Slope
59.87
Press ENTER key to next
mV/pH
1
▲
▲
Store the calibration results
1
▲
CAL
The calibration icon
will flash automatically
during calibration
→ YES
→ NO
Restore
1
▲
→ YES
→ NO
ORP Calibration
Calibration
→
Manual Calibration
Restore
→
1
▲
Manual Calibration
ORP Value :
Input Buffer :
0238mV
0238mV
1
▲
www.jensprima.com
Clean and dry the electrode and put it into the second point of the standard solution, enter the
current standard solution temperature (at MTC only), select the second point the standard solution,
press Enter to start cabration automatically, when stable, go to next page automatically