Parameter setup and measuring operation
42
LDM52 OEM, Release 003, Revision 000
6.8
TRI + TRO Trigger
6.8.1
Trigger function
The LDM52 OEM Trigger could be used as input or output.
1.
Trigger input / external trigger function:
External trigger signal will be sent
→
start of measurement DM in accordance with parame-
ter TRI.
Maximum trigger frequency: 1 Hz
2.
Trigger output / e.g. connection between 2 LDM52 OEM:
The output trigger signal of the 1. LDM52 OEM (parameterized with TRO) starts a single
measurement DM of the second LDM52 OEM (parameterized with TRO).
The frequency of the trigger output signal is determined by the output frequency of LUMOS.
When using the trigger function the output frequency has to be 1 Hz in maximum.
SA = 1
set MF < 1 Hz
SA > 1
set MF < SA x 1 Hz
Differences between trigger input and trigger output
Important is the parameter y of TRI and TRO.
TRI y > 0 / TRO y = 0
Trigger input
The measurement starts after an external trigger impulse.
TRI y = 0 / TRO y >0
Trigger output
LDM52 OEM sends a trigger impulse to the second device.
The parametrization of the trigger connection is carried out via a serial interface.
For the trigger function may only be activated TRI or TRO. A concurrent use of TRI and
TRO is not possible
→
output of warning information w1907