6
ideal EFFICIENCY (100%) for about four seconds before
continuing to 4mA setting.
3.
The [
¿
]
UP
key.
In
CALibration mode
where the user can change the settings, this
key will show the next possible option. If in decade number
format this key will
increment
the digit.
4.
The [
À
]
DOWN
key.
4a. In CALibration mode where the user can change the settings,
this key will show the next possible option. If in decade number
format this key will
decrement
the digit.
4b.During
pH Buffer Calibration
, the user can press this key to
repeat the buffer calibration.
5.
The [
OK
] key.
5a. In any mode where the user can change the settings, this key
will save the new settings. If no change has been made then
pressing this key will just move the user to the next setting.
5b. If the unit is password locked then continue pressing this key
will just move the user to view the next setting.
6
. The [
WASH
] key.
Pressing this key for about two seconds in any mode and the
unit is UNLOCKED will turn ON the WASH relay (Relay 5). If
the WASH relay is ON, The WASH relay will turn OFF after the
timer setting or the user may press this key to turn it OFF
immediately.