14
2.
Press
“Mode”
key until “mV” or “RmV” and “MAN” icons lit
up.
3.
Rinse electrode with distilled water and immerse it in sample to
be measured.Remove any air bubbles trapped around the
probe by shaking or stirring the probe.
4.
Allow sufficient time for the display to stabilize. The instrument
will display the mV (RmV) value of the sample.
G. Save, Recall and Delete Data
a. Saving readings to memory.
1. In the measure modes, press and hold
“Slope/Enter”
key
for 2 seconds to save data. The “ID” icon and number with
the corresponding site number will lit up for a brief moment
to indicate a successful data save.Saving is now complete.
2. If the “Full” icon is displayed, this means that all 750 data
saving sites are used up. No new data can be saved until
existing saved data are deleted.
b. Recalling readings from memory.
1. To recall saved data, press
“Mode”
key until the “Recall”
icon lit up. Press
"Slope/Enter"
key to go into “Recall”
mode.
2. Press the
“Up”
or
“Down”
keys to select the storage site
number.
3. Press
“Mode”
key to exit “Recall” mode.
c. Deleting data.
1.
Press
“Mode”
key until the “Delete one” or “Delete ALL” icon
shows up. Then press the
“Slope/Enter"
key to go into the
corresponding “Delete one ” or “Delete ALL” mode.
2.
In the “Delete ALL” mode, press the
“Slope/Enter”
and enters
the selection screen. Use the
“Up”
and
“Down”
key to select
yes/no for clearing all stored data
.
To clear all data, must
select yes and press the
“Slope/Enter”
key. “none” icon
shows up after data is completely deleted.
3.
In the “Delete one” mode, use
“Up”
and
“Down”
key to select
data to be deleted. Then press
“Slope/Enter”
key to delete.
Deletion is now complete. The next set of saved data will
automatically move up a slot in the storage site.