20
is then "locked" and will not respond to further changes
from the
pH electrode. The "locked" display is the pH value of the
sample at
the set sample temperature.
If "
OvEr/Undr
" is displayed, the pH value measured is out
of the
“16.00/-2.00 pH” range.
3.5 For samples that are inherently unstable, the instrument
will not autolock. Use the pH NON-AUTOLOCK mode for
measurements.
4.Measurement with manual temperature compensation in the pH
NON- AUTOLOCK mode.
4.1 Connect the pH electrode to the top of the instrument.
4.2 Press the [MODE] key for the LCD to display pH and for
the
AUTO annunciator to be off.
4.3 Rinse the pH electrode with distilled water and immerse it
in the
sample to be measured.
4.4 Set the instrument to the temperature of the sample as 2.2 of
pH
CALIBRATION section.
4.5 Allow sufficient time for the display to stabilize. The
instrument
Will display the pH value of the sample at the set sample
temperature.
If "
OvEr/Undr
" is displayed, the pH value measured is out
of the