OPERATION MANUAL
JENCO MODEL 6010 MICROCOMPUTER
BASED pH/mV/Temperature
PORTABLE METER
JENCO
ELECTRONICS, LTD.
MANUFACTURER OF PRECISION INSTRUMENTS
Страница 1: ...OPERATION MANUAL JENCO MODEL 6010 MICROCOMPUTER BASED pH mV Temperature PORTABLE METER JENCOELECTRONICS LTD MANUFACTURER OF PRECISION INSTRUMENTS ...
Страница 2: ...ATC Temp probe in pH mode 10 4 Calibration with MANual temperature compensation in the pH mode 12 pH MEASUREMENTS 13 1 Measurement with ATC Temp probe in the pH AUTOLOCK mode 13 2 Measurement with ATC Temp probe in the pH NON AUTOLOCK mode 14 3 Measurement with MANual temperature Compensation in the pH AUTOLOCK mode 15 4 Measurement with MANual temperature Compensation in the pH NON AUTOLOCK mode ...
Страница 3: ...s turned on again The front of the meter has a large LCD that displays the pH mV and temperature simultaneously along with the user prompting and mode indication annunciators The instrument prompts the user though the calibration and measurement procedures An AUTOLOCK feature is provided for both pH and mV measurements This enables the instrument to automatically sense the end point and lock the d...
Страница 4: ...hipment If any damage is found notify your JENCO representative immediately All packing materials should be saved until satisfactory operation is confirmed SPLASH RESISTANT The 6010 meter is housed in a water tighten case but should not be used underwater as the pH probe is not waterproof The water resistant feature is to prevent permanent damage to the instrument when accidentally drop into non c...
Страница 5: ...r appears on the LCD The instrument can operate within specifications for approximately one hour after LO BAT appears 1 Use a screw driver to remove the two screw and battery cover to expose the battery compartment Refer to Figure 1 2 Replace the 9V battery 3 Replace the battery cover and make sure to fasten the two screws for the water tight feature Figure 1 Battery compartment TURNING ON OFF THE...
Страница 6: ...This will be displayed if the unit is still waiting for a stable reading or end point sensing 2 BAT Low battery indicator 3 ATC MAN ATC indicator will displayed if an ATC probe is connected otherwise the MAN indicator will be displayed 4 STAND SLOPE This indicator will flash if the STAND or SLOPE is not yet calibrated respectively This indicator will remain lit up if the STAND or SLOPE is calibrat...
Страница 7: ...OFF The pH calibration values will not be erased when the instrument is turned off 2 MODE key The MODE key selects the parameters to be displayed Pressing the MODE key changes the display sequentially to display pH AUTOLOCK mV AUTOLOCK pH and mV The calibration values will not be affected by changing the display modes 3 STAND and SLOPE keys The STAND and SLOPE keys are used for dual point pH calib...
Страница 8: ...ibration values stored in the internal memory To prevent accidental pressing the CLEAR key Under normal use the key will not be activated unless press and held for 2 seconds The key will respond instantly only when error signal appears When the CLEAR key is pressed all segments of the LCD will be on After about 2 seconds the instrument will enter the pH AUTOLOCK mode The AUTOLOCK annunciator will ...
Страница 9: ... to press the CLEAR key 1 Calibration with an ATC Temp probe in the pH AUTOLOCK mode 1 1 Connect the combination pH electrode to the BNC connector and the ATC Temp probe to into the ATC Temp connector of the instrument The ATC annunciator on the LCD will be on Press the MODE key as necessary to go to pH AUTOLOCK mode 1 2 Rinse the pH electrode and ATC Temp probe in distilled water and immerse in p...
Страница 10: ... displayed the buffer temperature is out of the 0 to 60 range Bring the buffer temperature with in the range and repeat 1 2 It is not necessary to press the CLEAR key 1 4 Remove the pH electrode and ATC Temp probe from buffer 7 00 or 6 86 and rinse both in distilled water After rinsing immerse both in buffer 4 01 or 9 18 or 10 01 pH 1 5 Press the SLOPE key The SLOPE annunciator will stop flashing ...
Страница 11: ...e for the electrode and ATC Temp probe to stabilize before taking any action If Er 3 is displayed the buffer temperature is out of the 0 to 60 range Bring the buffer temperature within the range and repeat 1 2 It is not necessary to press the CLEAR key 2 Calibration with manual temperature compensation in the pH AUTOLOCK mode 2 1 Connect the pH electrode to the top of the instrument The MAN annunc...
Страница 12: ...ressed before the electrode and ATC Temp probe settle to within 1 5 pH of the buffer value Allow sufficient time for the electrode and ATC Temp probe to stabilize before taking any action 2 4 Remove the pH electrode from buffer 7 00 or 6 86 rinse with distilled water and immerse in buffer 4 01 or 9 18 or 10 01 Set the instrument to the temperature of the 2nd buffer as in 2 2 The displayed temperat...
Страница 13: ...e the electrode and the ATC Temp probe settle to within 30 of the buffer value Allow sufficient time for the electrode and ATC Temp probe to stabilize before taking any action 3 Calibration with ATC Temp probe in pH mode 3 1 Connect the pH electrode and the ATC Temp probe to the top of the instrument The ATC annunciator will be on Press the MODE key for the LCD to display pH and the AUTOLOCK annun...
Страница 14: ... stabilize before taking any action If Er 3 is displayed the buffer temperature is out of the 0 to 60 range Bring the buffer temperature with in the range and repeat 3 2 It is not necessary to press the CLEAR key 3 4 Remove the pH electrode and ATC Temp probe from buffer 7 00 or 6 86 Rinse them in distilled water and immerse in buffer 4 01 or 9 18 or 10 01 The instrument will display the temperatu...
Страница 15: ... within 30 of the buffer value Allow sufficient time for the electrode and ATC Temp probe to stabilize before taking any action If Er 3 is displayed the buffer temperature is out of the 0 to 60 rang Bring the buffer temperature within the range and repeat 1 2 It is not necessary to press the CLEAR key 4 Calibration with manual temperature compensation in the pH mode 4 1 Connect the pH electrode to...
Страница 16: ... within 1 5 pH of the buffer value Allow sufficient time for the electrode and ATC Temp probe to stabilize before taking any action 4 4 Remove the pH electrode from buffer 7 00 or 6 86 rinse with distilled water and immerse it in buffer 4 01 or 9 18 or 10 01 Set the instrument to the temperature of the 2nd buffer as in 4 2 The displayed temperature must be greater than 60 4 5 Allow sufficient time...
Страница 17: ...s pressed before the electrode settles to within 30 of the buffer value Allow sufficient time for the electrode to stabilize before taking any action pH MEASUREMENTS The STAND and SLOPE annunciators must be on This means that the instrument is dual point calibrated and is ready for measurements 1 Measurement with ATC Temp probe in the pH AUTOLOCK mode 1 1 Connect the pH electrode and ATC Temp prob...
Страница 18: ...he locked display is the pH value of the sample at the displayed sample temperature If OvEr Undr is displayed the pH value measured is out of the 16 00 2 00 pH range 1 5 For samples that are inherently unstable the instrument will not AUTOLOCK In this case use the pH NON AUTOLOCK mode for measurements 2 Measurement with ATC Temp probe in the pH NON AUTOLOCK mode 2 1 Connect the pH electrode and th...
Страница 19: ...for the LCD to indicate pH and for the AUTO annunciator to be on 3 3 Rinse the pH electrode with distilled water and immerse it in the sample to be measured Set the instrument to the temperature of the sample as in 2 2 of pH CALIBRATION section 3 4 Press the MEA EFF key the WAIT annunciator will start to flash This means that the instrument is waiting for a stable reading The display will track th...
Страница 20: ...anual temperature compensation in the pH NON AUTOLOCK mode 4 1 Connect the pH electrode to the top of the instrument 4 2 Press the MODE key for the LCD to display pH and for the AUTO annunciator to be off 4 3 Rinse the pH electrode with distilled water and immerse it in the sample to be measured 4 4 Set the instrument to the temperature of the sample as 2 2 of pH CALIBRATION section 4 5 Allow suff...
Страница 21: ...20 Once the temperature is brought within range Undr OvEr will disappear and the correct temperature reading will be displayed It is not necessary to press the CLEAR key mV MEASUREMENTS 1 Measurement in the mV AUTOLOCK mode 1 1 Connect the optional combination ORP electrode to the instrument in the BNC connector 1 2 Press the MODE key for the LCD to indicate mV and for the AUTO annunciator to be o...
Страница 22: ... 1250 2000 mV range The instrument will display the correct value once the input mV is brought within range It is not necessary to CLEAR the instrument 1 5 The ATC Temp probe can be used to measure the sample temperature as required 1 6 For samples that are inherently unstable the instrument will not autolock In this case use the mV NON AUTOLOCK mode for measurements 2 Measurement in the mV NON AU...
Страница 23: ... pH calibration buffers 4 01 6 86 7 00 9 18 and 10 01 are stored inside the instrument The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values TABLE 1 Temperature coefficient of the pH buffers 4 00 6 86 9 18 4 01 7 00 10 01 0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 2...
Страница 24: ...R DISPLAYS MAIN DISPLAY POSSIBLE CAUSE S Er 1 pH electrode offset is greater less than 1 5 pH pH electrode is faulty Er 2 pH electrode slope is off by more than 30 of the ideal slope Buffer 4 01 9 18 or 10 01 is not correct Er 3 Temperature is out of the 0 0 to 60 0 range SPECIFICATIONS Display Range Accuracy Resolution pH 2 00 to 16 00 pH 0 1 1 digit 0 01 pH ...
Страница 25: ...ut Impedance 1013 Ω Power Supply Batteries Adapter One 9V battery Calibration Back up EEPROM Audio Feedback All Touch Keys End Point Sensing Hold Yes LCD Size pH mV Temp 11mm 7 mm high LCD Ambient Temp Range 0 to 50 Relative Humidity up to 95 ATC Probe Thermistor 10kΩ Battery Life Alkaline 1000 Hours Dimensions W x D x H 86 mm x 241 mm x 72 mm Weight 430 grams Batteries included WARRANTY Jenco Ins...
Страница 26: ...repaired at no charge Make sure that the instrument is properly packaged and insured against possible damage in shipment Authorization must be obtained from a Jenco representative before returning items for any reason SHANGHAI JENCO ELECTRONICS LTD 18 Wang Dong Zhong Road Sijing Town Songjiang SHANGHAI CHINA TEL 86 021 5761 9599 FAX 86 021 5761 9598 E MAIL jenco public sta net cn Homepage http www...