JENCO 3177MB Скачать руководство пользователя страница 12

11

1. Standard solution of 1413 uS/cm at 25

: Accurately weight

out 0.746 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

2. Standard solution of 12.90 mS/cm at 25

: Accurately weight

out 7.4365 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

3. Standard solution of 111.9 mS/cm at 25

: Accurately weight

out 74.264 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

[Note:

You can store the unused portion of the standard

solution in a plastic container for six months but the air space

between the cap and the solution must be kept to an absolute

minimum. Storing the excess solution below 4

can increase

the storage life. If you have any doubt of the accuracy of the

stored solution, a fresh batch should be prepared.

]

APP NAVIGATION

A. App Functions

The 3177MB is a conductivity/salinity/TDS/temperature

Bluetooth benchtop meter with iOS and Android apps for

tablets. Functions of the App includes, graphing, guided

calibration procedures, data storage, data sharing and data

print.

Basic

functionality

of

a

conductivity/salinity/TDS/temperature benchtop meter such

as taking measurements, calibration and storing data can be

performed on the meter or on the paired tablets.

B. App Download And Installation

Download the companion App for the 3177MB from the Apple

Store or Google Play Store. Continue to the next step after

the correct App is installed to the tablets.

C. Connecting The Meter With The App

1. Prepare the 3177MB for Bluetooth connection

Press the

“Power/BLE”

key for 2 seconds to turn on the

3177MB device. Press the

"Mode/Clear"

key to go to the

measure mode. Press and release the

“Power/BLE”

Key

again to turn on the 3177MB Bluetooth connection. The

Bluetooth light on the 3177MB will fast blink and it is ready for

Bluetooth connection.

Содержание 3177MB

Страница 1: ...Operation Manual MODEL 3177MB Cond TDS Salt Temperature Based Bluetooth Benchtop Meter 3177MB...

Страница 2: ...splay 4 B Operational Keys Description 5 MODES OF THE METER 6 OPERATIONAL PROCEDURES 7 A Calibration 7 B Conductivity Salinity TDS Measurements 9 C Save Recall And Delete Data 10 D Preparing Conductiv...

Страница 3: ...icators The unit prompts the user through calibration and measurement procedures The unit is also equipped with a non volatile memory allowing the user to store 750 different sets of readings This uni...

Страница 4: ...fter unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position 3 Put the battery cover back on the meter Use a Ph...

Страница 5: ...ndicates Salinity measurement 3 uS cm Microsiemens indicates Conductivity measurement 9 CELL K Indicates conductivity cell constant value 4 CAL This will be displayed when the unit enters into the cal...

Страница 6: ...ternal memory In the Recall and Delete modes press Mode Clear key to exit Recall and Delete modes respectively 3 Power BLE Bluetooth Press and hold this key for 2 seconds to power on and shut off the...

Страница 7: ...ES and no for data all delete In the Delete One mode view to be deleted data and data site numbers by pressing these keys MODES OF THE METER This instrument has six modes as follow 1 Conductivity mode...

Страница 8: ...rature Coefficient Temperature Reference Probe Basic Cell Constant Standard Solutions Calibration and K Value Input To access these sections 1 Turn the unit on In the conductivity TDS salinity mode pr...

Страница 9: ...e is 1 91 To change the Temperature Coefficient use the Up or Down keys to adjust the value between 0 00 to 4 00 Press Enter key to save the new value and go to the next calibration parameter Temperat...

Страница 10: ...tant to a known value by pressing the Up or Down keys You can now input the K value from 70 to 130 of the probe basic cell constant Press the Enter key to exit calibration and return to normal operati...

Страница 11: ...a a Press Mode Clear key until the dEL onE or dEL ALL shows Then press the Enter key to go into the corresponding Delete one or Delete ALL mode b In the Delete ALL mode press the Enter and enters the...

Страница 12: ...prepared APP NAVIGATION A App Functions The 3177MB is a conductivity salinity TDS temperature Bluetooth benchtop meter with iOS and Android apps for tablets Functions of the App includes graphing gui...

Страница 13: ...7MB is connected to the tablet the device Bluetooth light will blink slowly After the App has been installed calibrate the 3177MB to ensure the best reading quality D Measurement Screen The Measure Sc...

Страница 14: ...ata From this screen tap the Calibrate button and follow the step by step instructions to calibrate the 3177MB Fig 2 G Settings Screen The Settings Screen displays the various settings that affects th...

Страница 15: ...rmation c From the saved reading detail screen tap the Notes field to add notes d Tap the button to save and exit the detail screen 6 Share and print saved readings a Tap the Saved tab in the navigati...

Страница 16: ...d Wait for the App to discover the 3177MB e Tap the Connect button I Notes 1 One meter can be paired to only one App at a time To put an already paired meter into pairing mode one must first close the...

Страница 17: ...olled environment This device and its antenna must not be co located or operating in conjunction with any other antenna or transmitter To comply with FCC RF exposure compliance requirements this grant...

Страница 18: ...t K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 10 0 to 120 0 1 Incorrect standard solution 2 Conductivity...

Страница 19: ...80 0 ppt 0 1 ppt 0 5 F S Temperature 10 0 to 120 0 C 0 1 C 0 2 C Conductivity calibration 1 point Reference Temperature 15 0 to 25 0 C default at 25 0 C Temperature Coefficient 0 0 to 4 0 default at...

Страница 20: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Отзывы: