![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 500](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616500.webp)
JD780A Series Analyzers
500
Using Cable and Antenna Analyzer
Connecting a cable
Connecting a cable for reflection, DTF, 1-port phase, and Smith
Chart measurements
Procedure
1.
Connect the one end of the port extension cable to the
Cable and Antenna Analyzer
Reflection/RF Out
port of the instrument.
2.
Connect the other end of the port extension cable to an antenna or feed line as illustrated in
Figure 251.
Measurement result is displayed on the screen.
Figure 251
Connection for reflection measurement
NOTE
After calibration, do not change the connection of the port extension cable or the
frequency setting. It may cause a measurement error.
Changing any frequency settings will automatically turn the calibration status to OFF on
the display screen, which requires re-calibration to proceed next measurement.
CAUTION
The maximum input power for the
Cable and Antenna Analyzer Reflection/RF Out
port
is +25 dBm. Do not connect this port directly to the output port of the system to prevent
degraded performance or malfunctioning.
WARNING
Do not attempt to connect the instrument to the antenna when there is a risk of lightning.
Electric shock may cause malfunction of or damage to the instrument.
Connecting a cable for cable loss (1 port) measurements
Procedure
1.
Connect the one end of the cable under test to the
Cable and Antenna Analyzer Reflection/RF
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...