background image

the gas conditioner to the analyser. It can be turned 
on/off via a switch on the front panel. In case of alarm 
condition the pump is locked automatically. 
 
 

Temperature display 

3.7.

To monitor the heat exchanger temperature, the instru-
ment is equipped with a digital display that shows the 
temperature or an error if the temperature limits are ex-
ceeded. 
 

Flow meter with needle valve 

3.8.

To adjust and display the gas flow, the instrument is 
equipped with a flow meter with integrated needle valve. 
 

CAUTION!

Never close the needle valve completely to avoid system 
damage. 
 
 

Condensate detection 

3.9.

The built-in electronic module monitors an eventually oc-
curring wet sample gas with the condensate sensor 
KW1. It turns of the power supply of the sample gas 
pump. Failure indication is done by colour changing of 
the green status LED “Condensate” to red. 

CAUTION!

 

In case that condensate detection occurs, the cause 
must be corrected immediately. To check or clean the 
sensor, open the nut and remove the sensor downwards. 
Equally dry the gas path downstream the heat ex-
changer. Check the unit for leaks before starting the op-
eration. 

CAUTION!

 

Condensate may contain hazardous substances! Wear 
appropriate protective clothing! 
 
 

Stainless steel sample gas inlet 

3.10.

A sample gas inlet of stainless steel SS316 is required 
to connect a heated sample hose. 

Temperature controller for heated sample line 

3.11.

To control a heated sample line the models with large 
housing can be equipped with a temperature controller.

Manual JCP-300 Series

aufbereitung zum nachgeschalteten Analysator. Sie 
kann durch den Kippschalter in Betrieb gesetzt werden 
und ist durch die optionale Temperatur- und oder Kon-
densatüberwachung verriegelt. 
 

Digitale Temperaturanzeige 

3.7.

Die Temperatur des Gaswärmetauschers und das Über- 
bzw. Unterschreiten der Grenzwerte wird auf der digita-
len Anzeige angezeigt. 
 
 

Durchflussmesser mit Nadelventil 

3.8.

Der Gasfluss wird durch einen in die Gerätefront einge-
bauten Durchflussmesser mit integriertem Nadelventil 
angezeigt und eingestellt. 

VORSICHT!

 

Das Nadelventil darf nie ganz geschlossen werden, 
sonst kann es zur Beschädigung oder Zerstörung von 
Systemteilen kommen. 
 

Kondensatüberwachung 

3.9.

Die eingebaute Auswerteelektronik überwacht den 
Messgasstrom auf Feuchtigkeit mittels Kondensatsen-
sor KW1. Dieser unterbricht die Spannungsversorgung 
der Messgaspumpe. Die Statusanzeige „Condensate“ 
wechselt von grün auf rot. 

VORSICHT!

 

Im Falle einer Kondensat Detektion, ist die Ursache zu 
beheben. Zur Kontrolle oder Reinigung ist der Konden-
satsensor nach Lösen der Überwurfmutter nach unten 
auszuziehen. Ebenso sind alle Gaswege nach dem Gas-
wärmetauscher zu trocknen. Vor Inbetriebnahme ist die 
Einheit auf Dichtheit zu prüfen. 

VORSICHT!

 

Kondensat kann gefährliche Substanzen enthalten! Ent-
sprechende Schutzkleidung tragen! 
 

Gaseingang aus Edelstahl 1.4401 

3.10.

Ein Gaseingang aus Edelstahl 1.4401 ist Voraussetzung 
für den Anschluss einer beheizten Messgasleitung. 
 

Temperaturregler für beheizte Messgasleitung  

3.11.

Über den Temperaturregler kann die beheizte Messgas-
leitung geregelt werden.

BA_DE_JCP-300_v1.0

          

–––––––––––––––––––––––––– [ 7 ] ––––––––––––––––––––––––––

Содержание JCP-300

Страница 1: ...JCT ANALYSENTECHNIK GMBH BEDIENUNGSANLEITUNG OPERATING MANUAL 3 20 JCP 300...

Страница 2: ...drain 12 6 5 Leakage test 14 6 6 Electrical connections 14 6 7 Connector plug terminal strip 14 7 Start up 15 8 Operation 16 9 Temperature controller heated sample line 16 9 1 General 16 9 2 Display o...

Страница 3: ...d Die JCT Analysentechnik GmbH haftet nicht bei eigen m chtigen nderungen des Ger tes oder f r deren un sachgem e Handhabung oder Verwendung Wenn anzunehmen ist dass ein gefahrloser Betrieb des Ger te...

Страница 4: ...ystemen bestimmt Beachten Sie die Angaben in den technischen Spezifikationen hinsichtlich Umge bungs und Versorgungsbedingungen sowie zul ssige Druck und Temperaturgrenzen Dieses Ger t darf nicht in e...

Страница 5: ...aus einem zylindri schen Hohlk rper Die Gaswege im W rmetauscher sind so gestaltet dass das feuchte Gas in Kontakt mit der gek hlten Oberfl che tritt Aufgrund der Gasf hrung bilden sich sowohl lamina...

Страница 6: ...s pump 3 6 The sample gas pump forwards the sample gas through erzeugter W rmetransport entzieht ber den Gasw r metauscher dem zugef hrten Messgas die Energie und f hrt diese durch die K hlrippen ab V...

Страница 7: ...ng can be equipped with a temperature controller Manual JCP 300 Series aufbereitung zum nachgeschalteten Analysator Sie kann durch den Kippschalter in Betrieb gesetzt werden und ist durch die optional...

Страница 8: ...with trolley function polypropylene yellow Gas wetted parts depending on heat exchanger aluminium coated PVDF SS316Ti FFKM Duran glass Dead volume per gas path 67 ml Connection sample gas inlet SS316...

Страница 9: ...K1233009A Rollentr ger Kondensatpumpe K1233011 Laufband 17 00002 Filterelement Glasfaser 5 Stk 17 00003 Filterelement PTFE 3 Stk 17 90001 O Ring Dichtung Filtergeh use VPE 3 Stk 12 90431 Verschlei te...

Страница 10: ...le gas inlet temperature 140 C Ersatzteile K1004600 Ventilator 230 VAC K1004605 Ventilator 115 VAC K1204323 Peltier Regelelektronik JPCU 1 K1204360 Kondensatw chter Elektronik KW 2 17 04000 Kondensats...

Страница 11: ...sensor KW 1 Option 5 Condensate sensor KW 1 option Messgaspumpe Option 6 Sample gas pump option Feinstaubfilter Option 7 Fine dust filter option Durchflussmesser mit Nadelventil 8 Flow meter with need...

Страница 12: ...ist das Ger t auf allf llige Trans portsch den zu untersuchen Wurde ein Schaden fest gestellt sind unverz glich die verantwortliche Spedition und der H ndler zu benachrichtigen Es ist zu berpr fen ob...

Страница 13: ...e line Filterelement 3 Filter element Kondensatsensor KW 1 4 Condensate sensor KW 1 Durchflussmesser mit Nadelventil 5 Flow meter with needle valve Temperaturregler f r beheizte Messgasleitung 6 Tempe...

Страница 14: ...g 6 5 VORSICHT Nach dem Anschluss ist die Dichtheit aller Schlauchver bindungen zu pr fen Elektrischer Anschluss 6 6 rtliche Netzspannung Netzfrequenz und Leis tungsaufnahme mit den Angaben am Typensc...

Страница 15: ...schl sse VORSICHT Betriebsspannung und max zul ssige Stromaufnahme von 8A beachten Heizleitung nur spannungslos an schlie en VORSICHT Dieses Ger t wird mit Netzspannung betrieben Beim Betrieb stehen z...

Страница 16: ...Manual JCP 300 Series 7 Energieversorgung herstellen Das Ger t beginnt zu k hlen Der Ventilator muss drehen und die Peltier Status LED leuchtet f r ca 15 min rot und die optionale Temperaturanzeige z...

Страница 17: ...den am Ger t f hren BA_DE_JCP 300_v1 0 17 Betriebszustand Status Status LED Condensate Gr n 1 green 1 Status LED Condensate Rot 1 red 1 Status LED Condensate blinkt 1 blinking 1 Status LED Peltier Gr...

Страница 18: ...arameter Wert werden angezeigt Dieser Modus erlaubt eine Einstellung von SV und des Alarmwertes 2 Dr cken der Block Taste um Setup Modus zu akti vieren Der Setup Modus erm glicht die Einstellung der P...

Страница 19: ...tions Remove probe from process flange Manual JCP 300 Series 5 Einstellung des Temperatur Sensors Zwei Varianten von Temperatursensoren werden ge meinsam mit dem JPXE4 verwendet 2x dr cken um Setup Ch...

Страница 20: ...iliar with the dangers involved Manual JCP 300 Series Anschluss des Kondensatablaufs trennen Fachgerechte Aufbewahrung bzw Entsorgung Recycling Das Ger t enth lt Bauteile die wiederverwertet werden k...

Страница 21: ...geltenden Umwelt Sicherheits und tech nischen Richtlinien zu erfolgen Reinigen der K hllamellen 12 1 Zum Reinigen der K hllamellen den r ckw rtigen De ckel abnehmen und Lamellen mit Druckluft freibla...

Страница 22: ...clockwise ro tation Schritt 2 Step 2 Die beiden Kondensat pumpen schl uche von den Fittings abziehen Pull off both condensate pump tubes from fittings Schritt 3 Step 3 Verschlussriegel im Uhrzeigersin...

Страница 23: ...bands am Pumpenkopf montie ren und beide Enden einschnappen Mount the tubing cover on the pump head place both end parts in the rail until they snap in Schritt 6 Step 6 Verschlussriegel gegen den Uhrz...

Страница 24: ...e two screws for pump head fixing and pull pump head with pulley holder off Schritt 2 Step 2 Pumpkopf mit neuem Rollentr ger auf Achse aufschieben Wellenschulter muss nach vorne zeigen Achten Sie dara...

Страница 25: ...t gefroren Umgebungstemperatur 1 C ggf Betriebsraum beheizen Kondensat alarm Ursache beseitigen siehe feuchtes Messgas Kondensatsensor durch L sen der ber wurfmutter nach unten abziehen reinigen und t...

Страница 26: ...DIN 34 Schutzvermerk nach DIN 34 beachten The technical data and design disclosed Analysentechnik GmbH and or contain propriertary rights of others and are not to be use sap 04 03 2020 Benennung P340...

Страница 27: ...tokoll 15 BA_DE_JCP 300_v1 0 27 T tigkeit Activity durchgef hrt von performed by Datum Date Unterschrift Signature K ltemittelverfl ssiger reinigen Condenser cleaned Pumpenschl uche der Kondensat pump...

Страница 28: ...AATWWXXX IBAN AT25 2011 1829 4301 5600 UNICREDIT BANK AUSTRIA AG BIC BKAUATWW IBAN AT89 1100 0039 9372 2200 UID Nr VAT No ATU20559900 Kreisgericht Wr Neustadt FN 119682 d Gesch ftsf hrer CEO G nter Zi...

Отзывы: