JCM Part No. 960-000076R_Rev. 1
© 2007, JCM-American Corporation
6 - 4
Section 6
World Bill Acceptor (WBA®-1x/2x -SS & SS2)
Calibration and Testing
10. At this point re-insert the White Calibration
Paper and click on the
“OK”
Screen Button to
re-start the “Adjusting with White Reference
Paper” calibration procedure again.
11. Repeat Steps 5 through 9 three to five times
until the Figure 6-14 Screen appears.
Calibration with White and Black Paper is
now complete. Remove the final calibration
paper and close the Acceptor Head
(See Figure 6-15 a). Make sure that the tabs
on both side of the Acceptor Head are once
again firmly locked in place (See Figure 6-
15 b).
12. Click on the
“OK”
Screen Button to start the
paperless calibration procedure (i.e., no
calibration paper inserted).
When calibration begins the Figure 6-16
Screen will appear.
13. When the paperless calibration process is
complete, the Figure 6-17 Screen will
appear.
14. Insert the MAG Head Test Board into the
Acceptor and find the location where the
second line of MAG Head Test Board just
aligns above the Acceptor Head Roller
located on the lower tray (See Figure 6-18).
Make sure that the tabs on both side of the
Acceptor Head are once again firmly locked
in place.
15. Click on the
“OK”
Screen Button to start the
MAG Head Tester Calibration procedure.
The Figure 6-19 Screen will appear showing
the current value
(a)
and the peak value
(b)
detected after the MAG Board has been
inserted.
Figure 6-13
White Re-Calibration Request Screen
Figure 6-14
Remove Calibration Paper Screen
Figure 6-15
Removing Final Calibration Paper
a
b
Figure 6-16
Paperless Calibration Screen
Figure 6-17
Magnetic Test Calibration Screen
NOTE: Closing the Acceptor Head once
the calibration process starts will result in
an erroneous reading!
Figure 6-18
Aligning the MAG Head Calibration
Test Board