© 2016 JCM American Corporation
1 8
November, 2016
Lecture Notes
12.When the
“Write EEPROM Succeeded”
prompt appears, click the
“OK”
screen
button to close the screen. Then click the
Close
box
b
) to close the
Sensor Calibration Screen and complete the Calibration process.
W
HITE
L
EVEL
T
EST
To perform the DBV-500 White Level Test procedure:
1. Disconnect power from the DBV-500 Unit.
2. Set DIP Switch Block #1, Switch # 8 to ON.
3. Restore power to the DBV-500 Unit.
4. Connect the USB Type-A to Mini-B cable between your PC’s USB port and the
USB port located on the DBV-500 Unit.
5. Launch the “JCM Tool Suite Standard Edition” software application.
6. Click the
Service Mode
drop-down menu, then click
Sensor Adjustment.
7. Confirm that the DBV-500 Calibration Tool is running, as shown in Figure 11.
8. Click the
White Level Test
function screen button (Figure 11
a
).
The White Level Test screen appears momentarily (refer to Figure 12).
NOTE: The DBV-500 Unit’s Front Bezel LED will flash
Green
.
a
Figure 11
DBV-500 Calibration Tool