JB INDUSTRIES 

 •  DV-22N Operating Manual  •  630.851.9444  •  [email protected]

Display Resolution (HI RES) 

The Display Resolution menu determines whether pressure will be shown at high resolution. High resolution 

means that pressure is shown as-is, in one unit increments. When the setting is OFF, the reading display is 

rounded. 
Available settings are: ON, OFF. 
Factory default setting is “OFF”. 
From the HI RES sub menu:
1.  Press the ENTER button to scroll through available options. 
2.   To select and store an alternate option as a new setting, press and hold the ENTER button for three seconds 

on the selected option. Unit will display “STORED” and save your selection.

3.   Press POWER button once to cancel and return to main menu. Press POWER button twice to exit and return 

to normal operation mode. 

Note: Even if high resolution display is selected ON, pressure values above 1000 microns will be rounded off 

according to the current display pressure.

Auto Shutdown (AU-0FF) 

To conserve battery life, the DV-22N will power off automatically after 30 minutes of inactivity. 
Available settings are: OFF, 10, 15, 30, 45, 60, or 90 minutes. 
Factory default setting is “30” minutes. 
From the AU--OFF sub menu:
1.  Press the ENTER button to scroll through available options. 
2.   To select and store an alternate option as a new setting, press and hold the ENTER button for three seconds 

on the selected option. Unit will display “STORED” and save your selection.

3.   Press POWER button once to cancel and return to main menu. Press POWER button twice to exit and return 

to normal operation mode. 

Note: Even if auto-shutdown has been deactivated, a low battery condition will power off the DV-22N.

Connection Instructions

It is recommended to connect the DV-22N gauge hose directly to either the high or low side of the system for 

the most accurate reading. 
To check for a leak using the vacuum pump blank-off valve: a metal hose, o-ring coupler or copper tubing must 

be used. Standard hoses permeate and will not hold a vacuum. The gauge reading will increase naturally due 

to equalization of pressure. Reading should hold for a minimum of five minutes—a continual increase may 

indicate a leak within the system.

MAINTENANCE

Cleaning Vacuum Sensor

1.  Fill brass sensor fitting with a few drops of isopropyl alcohol.
2.  Cover the opening with a finger and gently shake to allow alcohol to rinse the sensor.
3.  Pour out alcohol and allow sensor to dry for 20 minutes before using gauge. 

Содержание DV-22N

Страница 1: ...OPERATINGMANUAL DIGITAL VACUUM GAUGE dv 22N ...

Страница 2: ... C Power Supply Three AA batteries Dimensions DV 22N 9 length x 3 25 width x 1 5 depth DV 24N 10 length x 8 25 width x 2 75 depth Shipping Weight DV 22N 86 lbs 39 kg DV 24N 1 6 lbs 73 kg Warranty 12 months FEATURES Power Button Battery Status Indicator Display Screen 1 4 SAE Female Flare Fitting Hook Enter Button Battery Cover back Unit Selection Audible and Visual Alarm ...

Страница 3: ... an additional three seconds to store and enable the selected alarm setting Press POWER button twice to exit and return to main display screen and normal operation mode To silence an active alarm press any button to disable DisplayScreenStartUp Press and hold POWER button for two seconds to power gauge on If unit display does not light up check for correct battery installation Startup display will...

Страница 4: ...the unit will beep once per minute to alert the low battery condition When battery voltage falls below 2 1 volts the shutdown timer is activated If voltage stays below 2 1 volts for thirty seconds the DV 22N will beep once and power off BatteryReplacement If unit display does not light up after holding POWER button for two seconds check for correct battery installation 1 Remove battery cover from ...

Страница 5: ... press the POWER button once to return to the main menu 6 To exit the main menu and return to normal operation mode press the POWER button again Note When scrolling through menu settings the POWER button acts as the cancel exit button UnitSelection UNIT The available unit selections are Microns PSIA inHg Pascals Torr mTorr and mBar Factory default setting is MICRONS When the UNIT menu is active a ...

Страница 6: ...n to normal operation mode BackLightPersistence B LITE Back Light Persistence determines the number of seconds that the back light will stay on after the last button activity Available settings are OFF 10 15 30 45 60 and 90 seconds Factory default setting is 30 seconds From the B LITE sub menu 1 Press the ENTER button to scroll through available options 2 To select and store an alternate option as...

Страница 7: ...y default setting is 30 minutes From the AU OFF sub menu 1 Press the ENTER button to scroll through available options 2 To select and store an alternate option as a new setting press and hold the ENTER button for three seconds on the selected option Unit will display STORED and save your selection 3 Press POWER button once to cancel and return to main menu Press POWER button twice to exit and retu...

Страница 8: ...ndustries RGA _________ 601 N Farnsworth Ave Aurora IL 60505 630 851 9444 Tel 630 851 9448 Fax WARRANTY DV 22N Digital Vacuum Gauges are warrantied against defects in materials and workmanship for 12 months from date of purchase JB products are guaranteed when used in accordance with our guidelines and recommendations Warranty is limited to the repair replacement or credit at invoice price our opt...

Отзывы: