Jarven Health Care OptiCell SMART N Series Скачать руководство пользователя страница 8

8

FEATURES, OPTICELL SMART

OptiCell SMART air mattress systems are designed to enable individual cells to be disconnected 
simply while a patient is lying on the mattress. The mattresses have three different types of 
cells with different functions.

•   Main  cells

 - Continuously inflated air cells used primarily under the head.

•   Standard  cells

 - dual chamber cells where some chambers are continuously inflated 

and others have alternating air pressure.

•   Ventilation  cells

 - act like standard cells, but with ventilation holes that blow small 

amounts of air upwards into the cover. This is to create air circulation that helps keep the 
top of the cover dry and reduce the risk of damage. 
By connecting in an additional ventilating cell, an airier feeling can be created if the user 
says the cover feels hot in a certain area.

Individual cells can be disconnected simply by:

 

•   Removing the air hose from the cell concerned by pressing the button on the air hose 

connector.

•  Disconnecting both buttons on both sides of the cell.

AIR HOSES FROM THE PUMP.

The OptiCell SMART has a connector with three hoses from the pump to the mattress. 
The intermediate hose (red), is used to inflate the main cells under the patient’s head and to inflate 
the lower cell chambers in the other cells.  This is to ensure there is always air in the mattress 
irrespective of the pump setting or if the pump were to stop working 

(power outage or similar)

The two other hoses supply either alternating or static pressure to the standard and ventilation 
cells. See section “Air hose connection” (page 10) for further information.

THE DIFFERENT PUMP SETTINGS.

By changing the settings, the mattress can be used in either static or alternating pressure 
mode with different degrees of air pressure. See Overview Pump, page 15.

Содержание OptiCell SMART N Series

Страница 1: ...1 USER INSTRUCTIONS OptiCell SMART air pressure mattress mattress series N R and T...

Страница 2: ...such assurance will be provided for at least ten years QUALITY ASSURANCE Our quality assurance and customer focused approach is based on listening to the needs and expectations of our customers and d...

Страница 3: ...6 Expected useful life 6 Fire Safety 6 Intended use 7 Features OptiCell SMART 8 Before using it 9 Handling 10 Air hose connection 11 Hand test 11 Overview 4 NX Pump OC55 12 Specifications 13 Maintenan...

Страница 4: ...ducts accordingly Please contact us if you have any questions To avoid the risk of damage these products should not come in contact with heat sources or fire such as lit cigarettes hot lights or simil...

Страница 5: ...uctions contain information for general use maintenance and improved safety It is important to read these user instructions before using the product These user instructions should be made available to...

Страница 6: ...d rinsed clean with warm water max 70o C or cleaned with a disinfectant agent The Low Air Loss function allows air to be released from the cells and is also waterproof This together with the elasticat...

Страница 7: ...cm 200 kg Single cell Bed SMART T7 18 cm 230 kg Single cell Combi MATTRESS SERIES N TPU CELLS MATTRESS THICKNESS MAX LOAD CELL TYPE MATTRESS TYPE SMART R6 15 cm 210 kg Cell on Cell Replacement SMART...

Страница 8: ...feels hot in a certain area Individual cells can be disconnected simply by Removing the air hose from the cell concerned by pressing the button on the air hose connector Disconnecting both buttons on...

Страница 9: ...relation to the air pressure in the mattress and the weight of the patient Bear in mind that a tall user can weigh as much as a short person but will need different pressure This must always be caref...

Страница 10: ...ontrol panel see page 12 NOTE THAT the correct therapeutic effect is achieved in alternating mode Static air pressure constant low pressure should only be used when this is necessary LOW PRESSURE ALAR...

Страница 11: ...hes that hold the connector in place To minimise the risk of air leaks at the connector check That the two catches are not damaged and still retain their resilience That the black O rings on the three...

Страница 12: ...revious settings after 30 minutes The buttons on the panel are locked to prevent unauthorised adjustment of the pump operating mode To unlock the panel press and hold in this button for 5 seconds Sele...

Страница 13: ...30 minutes SYSTEM SPECIFICATIONS Classification Class II IPXO No AP APG Type BF Applicable part Mattress Voltage AC 220 240 50 Hz 1A 17W Fuse 250 V T 1 A Environment criteria TEMPERATURE Use 10 35 C R...

Страница 14: ...oses for any damage or abnormal wear Check the air hose connection See page 11 INSPECTION OFTHE OPTITEX HYGIENE COVER Regularly check the cover for visible signs of damage or wear such as rips cracks...

Страница 15: ...ers bedding etc should be washed at min 70o C for at least 10 minutes with unperfumed detergent in accordance with standard SS EN 14065 2016 Textiles Laundry processed textiles Biocontamination contro...

Страница 16: ...and to lie still as usual It is quite common for a patient to raise their body thinking that this will make it easier for the care worker which would give incorrect test results Fold the sheet to the...

Страница 17: ...TTRESS OptiCell air mattresses can be deposited at a recycling station and sorted as Combustible Waste For more information please contact J rven Health Care J rven Health Care continuously seeks to i...

Страница 18: ...on is not relevant EXPECTED USEFUL LIFE Under normal conditions and normal care and maintenance our OptiCell systems have an expected useful life of 5 years PRODUCT RETURNS AND CLAIMS Claims must be s...

Страница 19: ...e of manufacturer The product is a medical device product 70 C MAX 95 C Machine washable at 95 degrees Machine washable at 70 degrees max 95 degrees Do not use chlorine based cleaning agent Can be tum...

Страница 20: ...RO and Gentle 90 cm 12 960402 004 Nylon PU cell ventilating OC 2 NX 2 PRO and Gentle 105 cm 12 960402 005 Nylon PU cell ventilating OC 2 NX 2 PRO and Gentle 120 cm 12 970103 008 OptiTex 4 hygiene cove...

Страница 21: ...21...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...iCell SMART User Instructions All trademarks are the property of their respective owners J rven AB reserves the right to update these user instructions if necessary and or to correct any typographical...

Отзывы: