Flexx two manual
Installation, operation and use
Jansen Medicars
V2 2016
Pagina 14
3.
When using the optional mounting pole, attached accessories can
influence the balance negatively and is at your own judgement and
responsibility.
4.
Setting up the cart for use, will also include the installation of the
(electrical) devices on the cart. Make sure they are attached firm and
correctly.
5.
After installation and functional tests, it is important to check the
leakage currents. The leakage current at the input should be
<500µA. This is defined by the transformer, so this specification will
be made when the transformer is functioning correctly. At the
secondary side of the transformer, the total leakage current is mainly
defined you the connected devices. Make sure the total leakage
current is <5mA.