Jandel HM21 Скачать руководство пользователя страница 5

practice, particularly above 1mV the accuracy is better than 0.5% so it would not 
be unusual to find that you are within 0.2% of the expected value. 

 

Press ‘REV’. The HM21 is now passing current in the reverse direction. Again you 
should  receive a reading of 100mV, but this time with a minus figure. 

Press the ‘

Ω / 

 button. For thin samples the most commonly used measurement is 

sheet resistance. This is measured in ohms per square. As the size of the square is 
not relevant to the measurement there are no units such as ‘square centimetre’ or 
‘square metre’. The formula for sheet resistance is quite complicated but can be 
resolved to 4.5324 x V/I. Therefore when using a 100 ohm resistor you can expect 
the reading to be approx 453.24 ohms/square. 

Press the ‘

Ω / 

 button again and the unit will once again show the reading in mV. 

Note that the ohms/square value is absolute and therefore does not have a positive 
or negative sign. 

 

Sometimes the voltage readings can be slightly offset. This offset can be eliminated 
by ‘zeroing’ the HM21. With the unit already in Standby, press the Standby button. 
The unit will display ‘Zeroing…’ and after a few seconds return to standby mode. 
When making measurements if you have different forward and reverse readings it 
is worth zeroing the HM2) to make sure that it is not being caused by an offset. If 
you still obtain readings which are very different in forward and reverse this can 
indicate a poor contact or invalid measurement site. 
 
The range of measurement in high resolution is 0-150mV. In Low resolution it is 0-
1250mV. Where practicable the high resolution should be used and a mV reading 
of 50-100mV is ideal. 
 
Use the ‘INC’ button to increase the current to 10mA. With a 100 ohm resistor we 
will expect a reading of 1000mV. Press ‘FWD’. You should encounter an error 
message as 1000mV is outside of the high resolution range. The unit will display 
‘Out of Range’. Press ‘Low’ and you should obtain the reading of 1000mV.  
 
With the unit back in Standby remove the resistor. Attempt to make a reading by 
pressing the ‘FWD’ key. This will cause an error and the message “Contact Limit” 
will be displayed. By removing the resistor we have made it so that the HM21 can’t 
drive the current –  you can therefore equate this to a four point probe being on a 
sample but the HM21 having a problem driving the current through. This could be 
because the sample is too insulating, because the current needles are not in proper 
contact with the sample, or simply that a lead is disconnected!  Using a lower 

Содержание HM21

Страница 1: ...to operate the unit directly from a PC and to output results as sheet resistance bulk resistivity or bulk resistivity for wafer values The HM21 enables current to be passed in a forward and reverse d...

Страница 2: ...et resistance in ohms square depending which function has been chosen Unpacking When you unpack the HM21 you should find 1 A power lead incorporating a 9V AC adapter 2 A set of four AA rechargeable ba...

Страница 3: ...increases the current by one level 10mA max When pressed the DEC button decreases the current by one level 100nA min When pressed the button changes the display from mV to ohms square Pressing the but...

Страница 4: ...ery Low The warning will first be noted at the higher current levels or when changing from Reverse to Standby The unit can be operated via the adapter whether the batteries are inserted or not The bat...

Страница 5: ...mode When making measurements if you have different forward and reverse readings it is worth zeroing the HM2 to make sure that it is not being caused by an offset If you still obtain readings which a...

Страница 6: ...than 0 5 where the DVM receives greater than 1mV For the mid ranges the accuracy is better than 0 3 The DVM has two ranges high sensitivity up to 150mV and low range up to 1 25V The unit is push butt...

Страница 7: ...p the unit topped up but can t be used to charge the unit when the batteries are very low The batteries must always be rechargeable AA Nickel Metal Hydride or similar type The unit will completely tur...

Страница 8: ...g the data into a file on the PC will need to be used At the most basic level data can be captured in hyperterminal Save it as a text file change the files extensions to csv and it can then be opened...

Отзывы: