(II)
JWL 6
3. Press
M-/SET
to enter the parameter setting.
4. Press
/MR
or
TARE/
to shift between the function parameters
5. Press
M+/SAVE
to save and return.
6. Press
ZERO/ESC
to exit without saving.
5-2 Description of Parameter Values
1.
P00:
Offset value
Displays the offset value and the keypad testing can be conducted
2.
P01:
Backlight mode
Off
:
No backlight
Auto
:
when weight value is stable or any key is pressed. But auto off after N seconds (N=2s, 5s,
10s, 20s, ever) with no action
Ever=
It is always on when the weights over 9e
On
:
Backlight on
3.
P02:
Auto-off
Off
:
Non power off
5
,
10
,
30
,
60
(minutes): Auto power off after keeping the scale unused for
5
,
10
,
30
,
60
(minutes)
with
“
------
”
appears
4.
P03:
Unit setting
Init
: Press key
Unit
to
select the default unit when powering on the scale
:
pcs,
斤
, lboz, g, kg,
final .(final=keep the final being used unit when power off )
Use:
Press key
Unit
to select the weighing unit.
on
:
Enable the unit
off
:
Disable the unit
5.
P04:
Zero range
d0
,
d1
,
d2
,
d3
,
d4
and
d5
. (d= scale division)
6.
P05:
Hold function
HoLd – 0
:
no hold function
HoLd – 1
:
Peak hold. Press
M-/SET
key to release, and press
+/PRINT/HI
key to
print out
the value.
HoLd – 2
:
Hold after stable. Press any key to release
HoLd – 3
:
Hold after stable. Release after moving away the article. The hold value is based
on the current value and its range could be set in sub menu. Accumulation hold function is