Electronic Scale   

 

 

 

 

 

Pr Fu 

External device 

PC/ SH-24 

 

Item 10, Page 7 

 

 

 

 

 

diV-Z 

Display range 

Zero0/1/2/3/4/5 

 

Item 11, Page 7 

 

 

1.  When the LCD display shows Count, press the

UNIT

key, the current internal value 

will be displayed. Press

 

UNIT

to return to LCD display

 

  Count. 

 

2.  When the LCD display shows      unit    , press

UNIT

  to enter into unit selection mode, 

the  screen  will  show      on    g    or      oFF  g  , 

“on”  means  to  use  the  current  unit 

displayed, 

“oFF” means not to use the current unit displayed. Press

SMPL

  to switch 

between

 

ON and OFF [to use the unit displayed select 

“on”, if you do not require the 

unit  displayed  select 

“oFF”].  Press

UNIT

to  select  from  13  kinds  of  weighing  units, 

after selecting the unit required, press

MODE

to return to LCD display      unit      . 

 

3.  When the LCD display shows      InitU      , press

UNIT

 

to enter initial unit selection 

mode.  LCD  display    InitU 

xx

    X  means  the  initial  unit  is  selected.  Press

UNIT

 

to 

select the unit required, then Press

MODE

to return to LCD display

 

    InitU. 

 

4.  When  the  LCD  display  shows      Auto    ,  press

UNIT

  to enter  the  Auto Power-Off 

selection. LCD display      XXX      X means Auto Power-off selection. Press 

UNIT

 

to 

select the auto power-off time, then press

MODE

to return to LCD display      Auto    . 

Au-no- 

means  No  auto  power-off, 

5- 

means  5  minutes, 

15-

means  15  minutes, 

30-

means 30 minutes, 

60-

means 60 minutes

 

(in case of instability, the auto power-off 

function will not work) 

 

5.  When  the  LCD  display  shows      bAud    ,  press 

UNIT

  to  enter  the  RS-232  Baud 

Rate  selection.  LCD  display        XXX      X  represents  the  RS-232  baud  rate.  Press

UNIT

to select the desired

 

RS-232 baud rate, then press

MODE

  to return to LCD 

display      bAud    . 

2400, 4800, 9600

 

 

6.  When  the  LCD  display  shows      ZEro    ,  press

UNIT

 

to  enter  auto  zero  range 

selection. LCD display      d X        X represents the auto zero range. Press

UNIT

to 

select  the  desired  auto  zero  range,  then  press

MODE

  to  return  to  LCD  display       

ZEro   

d0, 1, 2, 3, 4, 5

  (the higher the value is, the wider auto zero range is) 

 

7.  When  the  LCD  display  shows        FiL      ,  press

UNIT

  to enter  filter  setting  mode. 

LCD display      FiL X      X means the filter

 

level. Press

UNIT

  to select the desired 

filter level and press

MODE

 

to return to LCD display

 

        FiL      . 

FiL1, 2, 4, 8

 

(the higher the value is, the higher the filter level is) 

 

8.  When the LCD display shows        LigH    , press

UNIT

to enter backlight setting mode. 

Содержание Jadever SKY

Страница 1: ...utions 1 Descriptions of Keys 2 Functional Operation of Keys 2 Panel Illustration 3 Single Point Calibration of Weight 3 Linearity Calibration of Weight 4 Parameter Setting 5 Parameter Functional Description 6 Units Conversion Table 7 Cross reference Table of Full Loads 7 RS 232 Serial Interface 8 ...

Страница 2: ... Steps 1 In the upright position locate the shipping protection screw device on the left side of the scale 2 Before using the scale unscrew counterclockwise and remove the shipping protection screw 3 Before transporting the scale hold the shipping protection screw down inwards and tighten it clockwise Precautions 1 Please pay attention to the environment when performing weight calibration on the s...

Страница 3: ...ou can start weighing If you need backlight press SMPL key Counting Press MODE key the screen will show XXXX pcs then press SMLP key and the screen will show S 20pcs You may select sample size of 20 50 100 200 or 500 in turn by pressing SMLP in succession After selection put the sample on the weighing pan and start counting operation when you hear a beep sound Automatic averaging function is provi...

Страница 4: ...ingle Point Calibration of Weight Step 1 Power off while holding down MODE key and then press ON OFF to switch on the scale The screen will show the following Count Step 2 Press the MODE key the screen will show the following CAL Step 3 Press the UNIT key to start zero point calibration Make sure there is nothing on the weighing pan The screen will show the following ZEro Please wait for the scree...

Страница 5: ...de Linearity Calibration of Weight Step 1 Power off while holding down ZERO and TARE keys and then press ON OFF to switch on the scale The screen will show the following L CAL Step 2 Clear the weighing pan then press TARE and the screen will show the following zero point calibration on0 Step 3 Wait for a beep sound and the screen will show the following on1 Step 4 Place 1 3 of load on the weighing...

Страница 6: ...e display Item 1 Page 6 CAL Single point auto calibration Page 3 unit Weight unit selection g ct lb oz dr GN ozt dwt MM tl j tl T g m2 tl H t oz yd2 for plaiting purpose ON OFF Item 2 Page 6 InitU Initial weight unit selection g ct lb oz dr GN ozt dwt MM tl j tl T g m2 tl H t oz yd2 for plaiting purpose Item 3 Page 6 Auto Auto power off selection Au no 5 15 30 60 Item 4 Page 6 bAud Baud rate selec...

Страница 7: ...r Off selection LCD display XXX X means Auto Power off selection Press UNIT to select the auto power off time then press MODE to return to LCD display Auto Au no means No auto power off 5 means 5 minutes 15 means 15 minutes 30 means 30 minutes 60 means 60 minutes in case of instability the auto power off function will not work 5 When the LCD display shows bAud press UNIT to enter the RS 232 Baud R...

Страница 8: ...l PC SH 24 external dot matrix mini printer 11 When the LCD display shows diV Z press UNIT to enter external device selection mode LCD display ZEroX X represents display range Press UNIT to select the desired external device then press MODE to return to LCD display diV Z Zero0 1 2 3 4 5 means weight displayed when the weight exceeds the capacity 12 Before entering function setting press SMPL to en...

Страница 9: ...6X0 01 385 92X0 02 964 80X0 05 1929 6X0 1 Momme 40 012X0 002 80 025X0 005 160 05X0 01 400 12X0 02 800 25X0 05 Tael HK Jewelry 4 0088X0 0002 8 0175X0 0005 16 035X0 001 40 088X0 002 80 175X0 005 Tael Taiwan 4 0012X0 0002 8 0025X0 0005 16 005X0 001 40 012X0 002 80 025X0 005 Tael Hong Kong 3 9696X0 0002 7 9390X0 0005 15 878X0 001 39 696X0 002 79 390X0 005 Tola India 12 865X0 0005 25 728X0 001 51 456X0...

Страница 10: ...ART BIT DATA BIT STOP BIT DATA FORMAT S U T G W LF CR SP SP WEIGHT SP UNIT ST STABLE UT UNSTABLE GW GROSS WEIGHT SP SPACE Examples ST GW 150 00 g UT NW 80 00 ct PC Program 10 OPEN COM1 9600 N 8 1 RS DS LF AS 1 BASICA 20 INPUT 1 A 30 PRINT A 40 GOTO 20 50 END ...

Отзывы: