Chapter 5
– Setup Menu
58
FIGURE 51 -
R
G
B
MEASUREMENT SCREEN
2. The instrument will make continuous measurements until the
Abort
key is touched. Adjust the primaries
of the display until the vertical indicator bars for
R
,
G
and
B
are all at the center of the respective
horizontal colored bars and the % deviation (values to the right of the bars) reads 0.00.
Note that updated
Luminance, CCT
and
chromaticity
values are continuously displayed during the
calibration sequence.
3. Touch
Abort
to
stop the measurement.
4. Touch
Back
to exit this screen and return to the Measurement Setup Screen at the completion of the
calibration.
C
OLOR
R
ENDERING
I
NDEX
(CRI)
–
O
PTIONAL FOR
PR-7XX.
Color Rendering Index is a method of testing luminaires or other light sources for their capability to render the
color of an object being illuminated by that source with respect to a standard light source
– does the object look
the same under both sources. Fourteen reference colors are used to determine the similarities between the test
light and the standard with the first eight colors used to calculate the average CRI
– see Figure 52.
FIGURE 52 - COLORS USED TO DETERMINE AVERAGE CRI
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......