C I N E B R A T E User Manual
20
Command
Description
M
Measure command for the PR-653. Returned datum is (luminance in fL & cd/m2; x, y;
CCT).
Output Format
:
qqqqq,Y
f
.Y
f
Y
f
Y
f
e+ee,Y
c
.Y
c
Y
c
Y
c
,e+ee,x.xxxx,y.yyyy,CCCC[CRLF]
where
:
q
= Status code, 00000 for success or error status
Y
f
= Luminance in fL
Y
c
= Luminance in cd/m
2
e = Exponent
x = CIE 1931 x
y = CIE 1931 y
C = CCT in Kelvin
Output Example
:
00000,1.865e+01, 6.390e+01,0.4035,0.4202,2856
Q
Quit (exit) remote mode.
V
View last measurement code.
X
Sets LCD contrast level.
TABLE 3 - REMOTE CONTROL COMMAND SUMMARY
D
ATA
C
ODE
S
UMMARY
The following Table summarizes all data codes and returned values for the CINEBRATE. These commands are
used in conjunction with the “D” command.
Data Code
Description
110
status, Serial Number
Output Format
: qqqqq,nnnnnnnnCRLF
where
: nnnnnnnn = Instrument Serial Number
Output Example:
00000,36124505
111
status, Instrument Name
Output Format
: qqqqq,mmmmmmCRLF
where
: mmmmmm = Instrument Model
Output Example:
00000,PR-653
112
status, Number of Calibrated Accessories, Number of Calibrated Apertures
Output Format
: qqqqq, ac, ap CRLF
where
: ac = Accessories, ap = Apertures
Output Example:
00000, 1, 1