J.ASSY VISUM Monitor Скачать руководство пользователя страница 4

TECHNICAL INSTRUCTIONS

  |  

Visum Monitor 

  |  

  |  

J.Assy Agricultural 

 |  

TECHNICAL INSTRUCTIONS

7

6

Tweaking your system

Sensor’s parameters  

The default parameter from the system are: 0

SENSITIVITY = 400
TIME = 9

If the system is accusing a “no flow” when there is a flow 
going through the sensor you can change the sensitivity 
of the system to a lower value. If in this case you increase 
the sensitivity you are actually generating more problem 
to your system. 
 
If you want an alarm as soon as the flow goes down you 
can increase the sensitivity value, in that case a less 
intense flow will generate an alarm from the sensor. 
 
The time can be change when a faster alarm is needed, 
so if the goal is having a faster alarm you can set a lower 
value in the sensor. 

In order to avoid any kind of false alarms, the TIME should 
be increased and the SENSITIVITY should be decreased. 
If the SENSITIVITY is set too low it is possible that the 
residual flow (the flow when the meters are turned off but 
the air flow is on) stays above this level, so the sensors 
will take longer to send the “no flow” signal when the 
meter stops.

Monitor’s parameter 

The default parameter from the system are:

MANEUVER DELAY

: 5 

NUMBER OF SENSORS TO ENTER IN A MANEUVER 
STATE

: 8 sending no flow  (when more than 12 rows are 

communicating and listed in the monitor – check using 
function 

F1

PERCENT OF SENSORS TO ENTER IN A MANEUVER 
STATE

: 75% sending no flow (when less than 12 rows 

are communicating and listed in the monitor – check 
using function 

F1

PERCENT OF SENSORS TO EXIT A MANEUVER STATE: 
50% sending flow 
The first information to be clear is that the system works 
using just one logic to enter in a maneuver state. The 
monitor does not use both parameter (number of sensor 
to enter in a maneuver state OR percent of sensors to 
enter in a maneuver state). Which one is used depends 
on how many sensors are present to the monitor. 

ATTENTION:

• Any change in these parameters, even when 
small, can generate or increase the false alarm if 
done wrong. Be sure that the concept for each 
parameter is clear before changing any of this 
configuration.

TIP:

•  The system works for many fertilizer prescription, 
but the most testing to define the parameters was 
done using a 140 lbs./ha @5mph ( 3,5 in between 
rows)  . If your system is operating at these values 
or higher check other possibilities before changing 
the sensor parameters.
•  If the system is working with a small or low 
prescription rate range you may need to change 
the sensitivity parameter. 
•  If only one sensor or a small number of sensors 
are presenting problems check if there is any 
obstruction in the tubes or meters before changing 
the parameters.

Содержание VISUM Monitor

Страница 1: ...100 Yellow 25 Magenta 100 Black TECHNICAL MANUAL version 4 0...

Страница 2: ...norme s RSS sans licence d Industry Canada Son utilisation est soumise aux deux conditions suivantes 1 Cet appareil ne doit pas causer d interf rences et 2 Il doit accepter toutes interf rences re ues...

Страница 3: ...addressed to the monitor The PERCENT OF SENSORS TO ENTER IN A MANEUVER STATE which is related with a percentage of sensor that should send a failure message to the monitor so it can identify a maneuve...

Страница 4: ...STATE 8 sending no flow when more than 12 rows are communicating and listed in the monitor check using function F1 PERCENT OF SENSORS TO ENTER IN A MANEUVER STATE 75 sending no flow when less than 12...

Страница 5: ...will wait until 5 seconds maneuver delay for the second no flow message and that will go on until 6 sensors send a no flow message S3 send a no flow message Maneuver delay starts to count After3secon...

Страница 6: ...y starts count After 3 seconds S3 send a flow message The maneuver delay starts over After 2 seconds S7 and S1 send a flow message The maneuver delay starts over After 5 seconds there is no flow messa...

Страница 7: ...ituation not represent a problem with the monitoring system The change of any of the parameters can easily generate unusual behaviors from the system and any change should be done carefully If the use...

Страница 8: ...lizer sensor type tS Seed sensor type 6 Then the display will show t meaning TIME for 2 seconds 7 Select the desired TIME by pressing and confirm by pressing 8 Then the display will show SE meaning SE...

Страница 9: ...nsor function you could try to change the RF channel with function A4 1 Select function A4 2 The display will show CH for 2 seconds 3 Selectthedesiredchannelbypressing andconfirm by pressing When numb...

Страница 10: ...by pressing and confirm by pressing 4 Confirm by pressing A5 Self test This function runs a self test routine in the monitor and is used only at the factory for quality insurance purposes TIP In order...

Страница 11: ...if the cable is intact Check if the is properly connected to a 12V 24V power supply red positive black negative Check if the fuse is plugged and intact Access function F2 and change beep volume Acces...

Страница 12: ..._______________________ _______________________________________________ _______________________________________________ NOTES _______________________________________________ __________________________...

Страница 13: ...www assyag com info assyag com January 2019...

Отзывы: