Adjusting the Collection Schedule

Once the appropriate column has been selected and mounted to the AFC tower, the 
collection schedule must be set. 

1. 

From the home screen, you can adjust the count, volume, and buffer volume for 
collection. Press 

-/+

 and follow the on screen prompts to save your parameters.

Setting up the Carousel

1. 

Load the collection tubes onto the carousel with the lids facing inwards. Ensure 
you have loaded enough tubes for your collection protocol.  

2. 

Gently place the carousel onto the carousel plate. Ensure that the carousel alignment 
hole engages with the raised carousel alignment pin on the carousel plate.

Remember to label your tubes before placing them into the AFC carousel.

Flushing the Column 

1. 

The next screen will ask you to mount the column reservoir. Note: If using a qEV 
single, a column reservoir is not needed.

2. 

Press

 OK

 to continue and follow the on screen prompts. It is recommended to flush 

your column with fresh, filtered PBS buffer.

Sample Loading and Initiating Collection 

1. 

The AFC will ask you to have sufficient buffer available to run the column and to 
load the sample. 

2. 

Ensure that there is no residual buffer left on the top of the column, remove any 
excess with a pipette. Now add your sample.  
 
It is important to only load the sample at this point. Adding buffer at the same 
time or before the sample has fully entered the frit of the column will result in 
incorrect sample loading. 

3. 

Press 

OK

 to continue.The column will start running and the carousel will move to 

the buffer collection position. The lighting will change to 

red

 to indicate that the 

collection has started. 

4. 

When the sample has fully entered the upper frit of the column, top up the column/
reservoir with the correct amount of buffer to resume collection of the buffer volume. 

5. 

When the buffer volume has been collected, the carousel will move to position 1, 
collect the specified volume and then progressively move to collect each successive 
individual fraction. If more collection tubes are required than a fully loaded carousel, 
the user will be instructed when to remove the cover and carousel, replace the tube 
positions with fresh tubes and continue the run.