26
IMPORTANT
Use a waveform measuring instrument (such as
oscilloscope and recorder) with an input impedance of
1 M
Ω
or more.
The output of the device is internally terminated.
Accurate measurement is not possible with waveform
measuring instruments with an input resistance of
50
Ω
.
Inspecting the Device Before Use
2.1 Inspecting the Device Before Use
Check the device for any damage that may have occurred
during storage or shipping, and verify proper operation
before use. If you find any damage or failure, contact your
authorized IWATSU distributor or reseller.
See “Before sending back your device for repair” (p. 85).
Items to prepare
• Model SS-530/SS-531 Current Probe
• Model PS-54 Power Supply (available as an option)
• Waveform measuring instrument (such as oscilloscope
and recorder)
Connecting the SS-530/SS-531 to a IWATSU Oscilloscope
option with a power supply module for current probes (option)
installed allows the SS-530/SS-531 to operate without the
PS-54 Power Supply. For more information, contact your
authorized IWATSU distributor or reseller.