13
High Voltage Probe HV-P60A
5. Specifications
5.1 Instrument specifications
Maximum input voltage
DC
60 kV
Continuous wave (sine wave)
Refer to "Figure 7 Maximum input voltage
and frequency" on next page.
Single-shot pulse
8
0 KV (Standard pulse wave: 1×40 μs)
Frequency bandwidth
DC to 50 MHz
(
-3
d
B
)
Rise time
7 ns or less
[Note
]
Rise time Tr is a calculation value from
the next expression.
Tr
[
ns
]=
350 / Band width
[
MHz
]
Attenuation ratio
2000
:
1
±
5
%
Waveform distortion
±
5
%
(10
℃
to 35
℃
)
[Note
]
Ringing and Overshoot are excluded.
Input RC
R
1000 M
Ω
±
5
%
C
5 pF
±
2 pF
Output cable
Approx. 4 m
Connector type
BNC type
Input RC of oscilloscope that can be connected
R
1M
Ω
C
10 pF to 40 pF
Dimensions
Refer to 5.2 Appearance.
Weight
Probe body
Approx. 2.0 kg
[Note] Output cable and Matching box are
included.
Whole
Approx. 3.2 kg
[Note] Accessories are included.
Environmental condition
Operating temperature range
-10
℃
to
+
50
℃
Storage temperature range
-10
℃
to
+
50
℃