![Ivoclar Vivadent Wieland Zenotec Select Hybrid Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/ivoclar-vivadent/wieland-zenotec-select-hybrid/wieland-zenotec-select-hybrid_cleaning-instructions_2097665002.webp)
Cleaning the collet / spindle
Click the button “Re-
move tool“ in the CNC
module of the software.
1
Clean the collet using the
brush of the Maintenance Kit
and make sure that the seal-
ings in the collet do not fall off.
Clean the inner cone of the spindle
with the cleaning cone of the Main-
tenance Kit.
Insert the measuring pin into the
collet and hold it firmly with one
hand. Manually unscrew the collet
using the knurled screw of the Spin-
dle Maintenance Kit.
Open the front flap.
Cleaning of the exterior
Cleaning with a cloth and, if necessary,
with a cleaning agent.
Remove all the disc holders from the
machine and clean them.
Clean the interior of the material
changer using a vacuum cleaner
and a moist cloth. Use the clean-
ing function in the software for
that purpose.
Remove all material residues in the work-
ing chamber with the brush and / or vac-
uum cleaner.
The measuring sensor should be sepa-
rately cleaned using a brush.
Check if water has accumulated in the
water separator and empty the con-
tainer, if necessary, by opening the dis-
charging screw.
Do you have any questions?
Our service and support team is pleased
to give you advice either via telephone
or by e-mail.
CAD/CAM Support
+ 49 72 31 / 37 05 - 400
[email protected]
0
2
4
8
10
6
Wieland Technik GmbH & Co. KG, Lindenstraße 2, 75175 Pforzheim, Fon +497231 / 3705-0, [email protected], www.wieland-dental.de
Cleaning instructions
Zenotec select hybrid
Working chamber and disc holder
Material changer
Measuring sensor
Compressed air maintenance
Daily
Weekly
Monthly
6
5
4
3
Take a small quantity of the collet
grease from the Maintenance Kit
and apply it on the slopes of the
collet.
7
Insert the measuring pin into the
collet. Screw the collet into the
spindle until it stops using the
screw-in piece.
8
Click the button “Basic set-
tings“.
9
Confirm the prompt to open the
collet.
2
683284 / DE-EN / 2016-03-24 / Rev.001