FRONT PANEL DESCRIPTION / EN

10/15/10 Version No. 00

EN-5

(1). LCD display:

A. Main display: T1, T2 or T1-T2 reading.
B. Secondary display

:

 T1 or T2 reading and MAX, MIN,

AVG reading.

C. Time display

:

 Time display (100-hour clock) shows elapsed

time when MAX, MIN or AVG is on.

D. Auto power-off mark (

 

:

 

).

(2).  H  HOLD key: Press  H  HOLD key to freeze or

unfreeze the display reading.

(3). 

 Power key: Press 

 key to turn the meter on or off.

(4). MX/MN key:

 Press "MX/MN" key to step through the maximum,

minimum and average readings.

 Press "MX/MN" key for 2 seconds to exit MX/MN mode.

(5).  T1/T2/T1-T2  key:  Press  T1/T2/T1-T2  key  to  toggle

between  T1,  T2  or  T1-T2  in  the
main or secondary display.

(6). MEM key:

  Press  the  "MEM"  key  once  to  store  a  single  set  of

logged data in memory.

 Press the "MEM" key for 2 seconds to enter continuous

data-logging mode. Press again to exit this mode.

(7).  SET  key:  Press  the  "SET"  key  for  2  seconds  then

release it to enter to interval time setting for
continuous data-logging mode.

www.

.com

[email protected]

1.800.561.8187

Содержание TES-1316

Страница 1: ...Instruction Manual TES 1316 Dual Data Logger Thermometer FR EN IT DE ES www com information itm com 1 800 561 8187...

Страница 2: ...10 Version No 00 TABLE OF CONTENTS TITLE PAGE 1 INTRODUCTION 1 2 SPECIFICATIONS 2 3 FRONT PANEL DESCRIPTION 4 4 OPERATION INSTRUCTIONS 7 5 MAINTENANCE 13 6 RECALIBRATION PROCEDURE 14 www com informat...

Страница 3: ...al temperature scale of 1990 ITS 90 Read the following information carefully before attempting to operate or service the meter When servicing use only specified replacement parts Environment condition...

Страница 4: ...9 F S type 2 0 to 1767 0 C 35 to 1999 9 F Display effective resolution J K T E and N type 0 1 C F K R and S type 1 0 C F K 0 1 C F K only for reference Measurement accuracy J K T E and N type 0 05 of...

Страница 5: ...ontinuity data logging capacity 4100 sets 2 2 General specifications Power supply Qty 6 AAA batteries Battery life approx 80 hours Auto power off 30 minutes If no key is pressed Low battery indication...

Страница 6: ...3 FRONT PANEL DESCRIPTION T1 T2 14 10 12 13 1 5 4 2 3 7 9 8 15 6 11 C D A B 16 TYPE MX AVG MN INTERVAL min sec hour min T2 T1 T1 T2 C F K KJTERSN T1 KJTERSN TYPE hour min min sec INTERVAL T2 MX T1 T2...

Страница 7: ...r off 4 MX MN key Press MX MN key to step through the maximum minimum and average readings Press MX MN key for 2 seconds to exit MX MN mode 5 T1 T2 T1 T2 key Press T1 T2 T1 T2 key to toggle between T1...

Страница 8: ...ode 10 TYPE key Press TYPE key to select the thermocouple type K J E T R S or N 11 READ key Press READ key to show manual memory logged readings Press again to exit this mode 12 C F K key Press C F K...

Страница 9: ...thermometer Plug the thermocouple s into the thermocouple input s as required If no thermocouple is plugged into the selected input or the thermocouple is open circuit the appropriate display will sh...

Страница 10: ...n and min sec on the LCD display The elapsed time since entering reading mode or the time at which the MAX MIN or AVG value occurred will appear on the time display Press T1 T2 T1 T2 key to toggle bet...

Страница 11: ...sed time T2 Averaging reading plus averaging time TYPE min sec K MX MN AVG C T1 T2 present reading T1 T2 Maximum reading plus elapsed time T1 T2 Minimum reading plus elapsed time T1 T2 Averaging readi...

Страница 12: ...in memory will be cleared 4 4 To Trigger One by One Datalogging Press the MEM key once and one set of readings will be stored in memory The LCD display shows MEM and a memory location number 01 to 98...

Страница 13: ...splay will show INTERVAL MEM and interval time Press or key until the display shows the required logging interval 3 to 255 seconds and then press key to select Press the MEM key for 2 seconds to start...

Страница 14: ...nnot be read using the READ function to show the data the on display 4 6 How to disable auto power off function The meter will automatically turn off if no key press occurs for 30 minutes Press key to...

Страница 15: ...with a damp cloth and mild detergent Do not use abrasives or solvents Clean and dry as required 5 2 Battery Replacement When the LCD display shows BT the battery has insufficient power to support an...

Страница 16: ...IBRATION PROCEDURE The thermometer should be calibrated once a year to ensure its continued accuracy Contact RS Components for further details of calibration service The address is given at the end of...

Отзывы: