1. SAFETY PRECAUTIONS AND PROCEDURES 

This clamp complies with IEC/EN61010-1. For your own safety and in order to avoid 
damaging the instrument, you’re recommended to keep to the instructions contained in this 
manual and read carefully all the notes preceded by the symbol 

.

 

 

Take extreme care for the following conditions while measuring: 

  Do not measure voltage or current in humid or wet environment. 

  Do not use the meter in presence of explosive gas (material), combustible gas 

(material), steam or dust. 

  Insulate yourself from the object to be tested. 

  Do not touch exposed metal (conductive) parts such as test lead ends, sockets, fixing 

objects, circuits, etc. 

  If you detect anomalies of testing end (metal part) and attachment of the meter such as 

breakages, deformations, foreign substances, no display, etc., do not take any 
measurement. 

  Measuring voltage over 20V as it might cause human body electricity conduction. 

 

The followings symbols are used on meter and in this manual: 

 

 

Caution: refer to the instruction manual. An incorrect use may damage the tester 
or its components 

 

 

High Voltage ranger: electrical shock risk 

 

 

Double insulated instrument 

 

 

AC Voltage or Current 

 

 

DC Voltage or Current 

 

 

Ground reference 

 

1.1. PRELIMINARY 

  This apparatus has been designed for use in an environment of pollution degree 2. 

Indoor use. 

 It 

measures 

CURRENT

 and

 VOLTAGE

 on CAT IV 600V and CAT III 1000V plants. For 

measurement categories please see § 1.4. 

  You must comply with the usual safety regulations aimed at protecting you against the 

dangerous electric current and protecting the instrument against an incorrect operation. 

  Only the leads supplied with the instrument guarantee compliance with the safety 

standards. They must be in good conditions and they must be replaced, if necessary, 
with an identical model. 

  Do not test or connect to any circuit whose voltage or current exceeds the specified 

overload protection. 

  Make sure that the batteries are installed correctly. 

  Before connecting the test probes to the installation, check that the function selector is 

positioned on the required measurement. 

  Make sure that the LCD and the range indicator show the same as the function 

desired. 

 

www.

.com

[email protected]

1.800.561.8187

Содержание HT9021

Страница 1: ...Copyright HT ITALIA 2012 Release EN 1 04 19 04 2012 ENGLISH User manual www com information itm com 1 800 561 8187...

Страница 2: ...y 7 4 2 5 Hz key 7 4 2 6 PEAK REL key 7 4 3 Functions of rotary switch description 8 4 3 1 AC Voltage measurement 8 4 3 2 DC Voltage measurement 9 4 3 3 Resistance measurement 10 4 3 4 Continuity test...

Страница 3: ...on manual An incorrect use may damage the tester or its components High Voltage ranger electrical shock risk Double insulated instrument AC Voltage or Current DC Voltage or Current Ground reference 1...

Страница 4: ...ntrol and laboratory use Part 1 General requirements defines what measuring category usually called overvoltage category is On 6 7 4 Measuring circuits it says OMISSIS circuits are divided into the fo...

Страница 5: ...ults In presence of distorted waves the readings are different Mean value instruments provide only the value of the fundamental wave while True RMS instruments provide the value of the entire wave inc...

Страница 6: ...R SUPPLY The instrument is battery supplied One battery 9V NEDA 1604 IEC 6F22 JIS 006P is included in the package Battery life about 200 hours When batteries are low the symbol appears on the display...

Страница 7: ...DE MAX MIN Hz PEAK REL function keys 7 LCD display 8 V CAP Temp input jack 9 COM input jack Fig 1 Instrument description 4 1 2 Alignment marks Put the conductor within the jaws on intersection of the...

Страница 8: ...d on each measurement except for continuity test diode test capacitance test frequency test and duty cycle test This mode is disabled keeping pressed MAX MIN key or moving the rotary switch 4 2 5 Hz k...

Страница 9: ...sition 3 Insert the red test lead plug into V CAP Temp jack and the black test lead plug into COM jack Fig 3 4 Connect the two long ends of test leads to the desired circuit then reading will be displ...

Страница 10: ...Fig 4 3 Connect the two long ends of test leads to the desired circuit then reading will be displayed 4 When O L symbol is displayed the result is over the maximum value that the instrument is able t...

Страница 11: ...Insert the red test lead plug into V CAP Temp jack and the black test lead plug into COM jack Fig 5 3 Connect the two long ends of test leads to the desired circuit then reading will be displayed 4 W...

Страница 12: ...plug into COM jack and perform continuity test on the object under test see Fig 6 left side Buzzer emits sound if the measured resistance value is less about 60 4 Pushing MODE key and select diode te...

Страница 13: ...ug into V CAP Temp jack and the black test lead plug into COM jack Fig 7 4 Connect the two long ends of test leads to the desired circuit then reading will be displayed Bargraph is disabled in capacit...

Страница 14: ...n Temp position 2 Pushing MODE key and select the kind of measure C or K symbols are shown at display respectively for Celsius or Farheneit temperature measurements 3 Insert the type K bead probe in V...

Страница 15: ...hese values will not be added to the measured value On DC measurement the zeroing operation of residual magnetization current is essential to obtain accurate results 4 Put the conductor to be tested i...

Страница 16: ...A positions for frequency measurements with jaws 2 Pushing Hz key the Hz symbol for frequency measurement or the symbol for duty cycle measurement are shown at display 3 Insert the red test lead plug...

Страница 17: ...ly expert and trained technicians must perform this operation Remove the test leads or the conductor under test before replacing the battery 1 Rotate the switch on OFF 2 Remove the test leads or the o...

Страница 18: ...meter emits a bip bip sound on 1000V range for VAC 770V Resistance and Continuity test Autorange Range Resolution Accuracy Buzzer Overload protection 600 0 0 1 1 0 rdg 5dgt 60 600VDC ACrms 6 000k 0 00...

Страница 19: ...dg 5dgt 10A 60A 50A 600A 1000VACrms 999 9Hz 0 1Hz 9 999kHz 0 001kHz Duty Cycle Autorange Range Resolution Sensitivity Accuracy 0 5 99 0 0 1 15Vrms 1 2 rdg 2dgt Pulse width 100 s 100ms Pulse frequency...

Страница 20: ...0 F Storage humidity 80 RH This product conforms to the prescriptions of the European directive on low voltage 2006 95 EEC and to EMC directive 2004 108 EEC 6 3 STANDARD ACCESSORIES Test leads Code 44...

Страница 21: ...lts due to servicing carried out by a person not authorized by the company faults due to modifications made without explicit authorisation of our technical department faults due to adaptation to a par...

Отзывы: