4
INITiate Subsystem
Initiate commands initialize the trigger system. This moves the trigger system
from the "idle" state to the "wait-for- trigger" state; which enables the instrument
to receive triggers.
•
It takes a few milliseconds for the instrument to be ready to receive a trigger
signal after receiving the INITiate command.
•
If a trigger occurs before the trigger system is ready for it, the trigger will be
ignored.
•
Use ABORt commands to return the instrument to Idle.
INITiate[:IMMediate]:ACQuire
Initiates the measurement trigger system.
Group
INITiate
Syntax
INITiate[:IMMediate]:ACQuire
Arguments
None
Default Value
None
Returns
None
Example
INIT:ACQ
Copyright © Itech Electronic Co., Ltd.
36