Calibration Commands
Copyright © Itech Electronic Co., Ltd.
33
Parameters
P1|P2|P3|P4
CALibration:VOLTage [:DATA] {<numeric value>}
Return actual output voltage value of calibration point.
Order syntax
CALibration:VOLTage [:DATA] <NRf>
Parameters
<NRf>
Example
CAL:VOLT 30.0002V
CALibration:CURRent:LEVel
This order can set current calibration point. P1
、
P2 must be calibrated orderly.
Order syntax
CALibration:CURRent:LEVel <point>
Parameters
P1|P2
CALibration:CURRent [:DATA] {<numeric value>}
Return actual output current value to calibration point.
Order syntax
CALibration:CURRent[:DATA] <NRf>
Parameters
<NRf>
Examples
CAL:VOLT 3.0002A
CALibration:STRing
Set the calibration information.
Order syntax
CALibration:STRing <parameters>
Parameters
The longest length is a string including 24 letters i.e related information the
customer records during running the calibration process, such as the calibration