Operating manual: PHOTOMETER FM-1 14
5.
Technical Specifications
Basic technical data is also visible at the back side sticker of the photometer.
Sensor
silicon photodiode, V
λ
filter
Spectral range
human eye like
Spectral error
< 5 % to CIE
Full viewing angle
20° (as standard)
Measuring ranges
0-100; 0-1.000; 0-10.000 cd/m
2
(
0- 65; 0- 650; 0- 6.500 cd/m
2
)
(see calibration)
Accuracy
> 3 %
Dew & Temperature control
heater 25 W
Sensor-room temperature stability
±1
°C
Temperature range
-40
°C to +50
°C
Power supply
230 V AC ±15 %
Frequency
50-60 Hz
Max. control power
450 W
Power consumption
< 30 W
Fuse
F 2,5 A
(5x20 mm)
Casing
IP65 ; Aluminium / Stainless steel / ABS
Size
395 x 120 x 180 (220) mm
Weight
3,9 kg
Fixture
wall, X - Y adjustable fixture / pole
-optional
Communication port
RS-232/485 -
optional
/ 9600 bps
Analog outputs:
0 - 100/1.000/10.000 cd/m
2
(0 – 65/650/6.500 cd/m
2
)
0 -10 V ; min. loop resistance: 1000
Ω
4 -20 mA ; max. loop resistance: 650
Ω
Contactor 230VAC outputs:
Number of channels
4
Output voltage
230 V AC (common fuse 2,5 A)
Adjustable range
0 - 10.000 cd/m
2
(0 –6.500 cd/m
2
); auto ranging
Averaging interval
0 - 720 s
5. 1.
Environmental Conditions
Maximum environmental temperature at use is +50
°C; 24 hour average temperature
must not 35
°C. Lowest environmental temperature at use is –40
°C. Relative air
humidity must not exceed 50
% at maximum temperature +50
°C. Higher humidity is
allowed at lower temperature (like: 95
% at +20
°C).
Sea-level elevation must not exceed 2000 metres. When using above 1000 metres
consider lower air dielectricity and different cooling conditions.
Содержание Photometer FM-1
Страница 2: ......
Страница 21: ...Notes ...
Страница 22: ... ...
Страница 23: ...V1 rev 12 2012 ...
Страница 24: ...Litostrojska cesta 44 d SI 1000 Ljubljana T 386 1 514 18 00 F 386 1 514 18 04 info itaia si www itaia si ...