3. 

INSTALLATION AND ADJUSTMENT 

 

(See Figures 4 & 5) 

3.2 

With no d-c power applied, connect the + 24Vdc in to the center terminal of the feed-thru 

terminal 1.  DO NOT APPLY POWER. 

 

3.3 

Connect the RF output SMA jack to the SMA or BNC RF input of the acousto-optic deflector 

(or a 50

 RF load, if it is desired to measure the modulator RF output power). 

 

3.4 

Connect the amplitude control source to the the D-type connector. 

 

Connect the modulation signal ( 0,0 - 10.0V) to pin 8  

 

Connect the signal return (0V) to pin 3 

 

To begin, apply a constant voltage of 10.0V. 

 

3.5 

Connect the external Interlock of the acousto-optic modulator (if fitted) to the enable inputs on 

the 9-pin D-type connector.  If no interlock is fitted then these inputs MUST be connected 

together. 

 

Connect pin 9 to the floating contact 1 of the interlock switch. 

 

Connect pin 4 to floating contact 2 of the interlock switch. (Grounded at driver) 

 

The interlock connection becomes open circuit disabling the RF output, if the temperature of 

the modulator exceeds 30ºC or the internal driver temperature exceeds 70ºC.  An LED 

indicator illuminates when the Interlocks are closed and the RF is enabled. 

 

3.6 

If required apply a 12V CMOS compatible or open collector  driver to the Digital Gate input. 

 

Connect the tuning signal ( 12v maximum) to pin 7  

 

Connect the signal return (0V) to pin 2 

This input is optional and may be left NC (Not Connected). 

A closed contact on this input will disable the RF Output.  Switching response time 10msec. 

 

3.7. 

Connect the tuning voltage source to the the D-type connector  

 

Connect the tuning signal ( 0,0 -10.0V) to pin1  

 

Connect the signal return (0V) to pin 6 

 

3.8 

Adjustment of the RF output power is best done with amplifier connected to the acousto-optic 

modulator 

 

Содержание RFA920T Series

Страница 1: ...T 110 90 130MHz 10V Tuning Input 10V Analog Modulation Input RFA920T 150 125 175MHz 10V Tuning Input 10V Analog Modulation Input ISOMET CORP 5263 Port Royal Rd Springfield VA 22151 USA Tel 703 321 830...

Страница 2: ...ized varactor voltage controlled oscillator The output frequency is proportional to the input tuning voltage with a scale factor of approximately 2 4MHz volt depending on the model The start frequency...

Страница 3: ...e required voltage is 24Vdc at a current drain of approximately 2 8A The external power source should be regulated to 2 and the power supply ripple voltage should be less than 100mV for best results T...

Страница 4: ...the floating contact 1 of the interlock switch Connect pin 4 to floating contact 2 of the interlock switch Grounded at driver The interlock connection becomes open circuit disabling the RF output if...

Страница 5: ...f the D type connector This will operate the deflector at its mid scan position 3 14 Align the deflector head to insure that the incident light beam is centred in the active aperture of the deflector...

Страница 6: ...th a soft ball of brushed short fibres removed cotton slightly moistened with clean alcohol Before the alcohol has had time to dry on the surface wipe again with dry cotton in a smooth continuous stro...

Страница 7: ...Control signals 9 pin D type Tuning Voltage 0 10V sig pn1 rtn pn6 Amplitude Control 0 10V sig pn8 rtn pn3 Enable Interlock Control Low On sig pn9 rtn pn4 Digital Gate Input NC or High On sig pn7 sig...

Страница 8: ...mp RF Output PWR ADJ Offset Tuning Figure 2a Driver block diagram generic SKT DB9 5 9 4 8 3 7 2 6 1 3 1 2 Int GATE Optional No Connect or Open Drain FREQ TUNING 0 0 10 V 1Kohm MOD Amplitude Control 0...

Страница 9: ...9 ISOMET 1 0 1 RF Carrier 1 0 1 Modulation Signal Input 1 0 1 Amplitude Modulated RF Output Figure 3 Typical Modulation Waveforms...

Страница 10: ...face and in the plane of deflection is given by BRAGG fc 2 v The separation angle between the zeroth order and mid scan point of the first order is given by SEP fc v The first order scan angle is give...

Страница 11: ...If applicable Coolant circuit not shown for clarity Interlock Connector SMB Coax or 3 pin Snap Normally Closed Contacts Pins 1 2 Not connected Pin 3 Tune 0 10V Analog Mod 0 10V Gate O C Vdc 1 2 3 1st...

Отзывы: