ISOMAG ML 252 Скачать руководство пользователя страница 6

 
 

  

6

 

 

252_EN_IS_1_3_00.doc

 

 
 

 

The converter can be used in two different modes: 

 

with continuous sampling 

with sampling to preset unit of time. 

 

q

CONTINUOUS SAMPLING 

 (pic.1)  

(ENERGY SAVING OFF)

 

 

In this mode the converter make the measure in accordance with the classical diagram of the 

flow meter; the consumption of the system, with any diameter of the sensor is 0,05 W ; 

the 

life of battery is about 1 month (6 with 6 battery)  

The accuracy of the system is definite in those conditions. 

 

q

SAMPLING TO PRESET UNIT TIME 

( pic. 2 )  

( ENERGY SAVING ON )

 

 

This mode works sampling the range to intervals of preset time (see MEASURE menu, func. 

3.5); it allows a great saving of energy 

 

In this conditions the consumptions are: 

 

 
 
 
 
 

 
 
 

 
 
 
 
 
 
 
 
 
 

Sampling 

time (s) 

Battery life - n° 1 

battery (years) 

Battery life – “ N “  

batteries (years) 

 

0,7 

0,7 * N  

1.3 

1.3 * N 

2.1 

2.1 * N 

10 

2.7 

2.7 * N 

15 

3.0 

3.0 * N 

>= 30 

5.0 * N  

 

Whatever the results , 

the maximum time is  

limited to 10 years 

FLOW RATE

SAMPLE 

FREQUENCY

SAMPLE 
TIME

 

t

t

FLOW RATE

SAMPLE 

FREQUENCY

PIC 1 

PIC 2 

ATTENTION: 

the consumptions on the table are without optional IF2 cable connected to the 

converter. This device accelerate the consumption of the batteries even if the instrument is in standby 

mode

. Recommends to disconnect the IF2 cable from the converter after every his uses or 

switch off the converter by the dip-switches (see page 9 )

 

MEASURE/ CONSUMPTIONS  

 

Содержание ML 252

Страница 1: ...OPERATING AND INSTALLATION MANUAL CONVERTER ML 252...

Страница 2: ...___________________________________pag 9 q How to access at the instrument functions ________________________________________________________ pag 10 User interface ____________________________________...

Страница 3: ...ious notice The flow meter must be used for what it has been built for The improper use possible tampering of the instrument or parts of it and substitutions of any not original components makes the w...

Страница 4: ...4 252_EN_IS_1_3_00 doc 125 112 97 OVERALL DIMENSIONS...

Страница 5: ...pply frequency Max power Max Current LITIUM BATTERY 3 6 V 16 5 A h INPUT OUTPUT ISOLATION q Input output are insulated up to 500V q Port RS 232 NON is not insulated ENVIRONMENTAL CONDITIONS OF USE q T...

Страница 6: ...als of preset time see MEASURE menu func 3 5 it allows a great saving of energy In this conditions the consumptions are Sampling time s Battery life n 1 battery years Battery life N batteries years 1...

Страница 7: ...lated output Opto MOS q Maximum switching voltage 40 Vdc q Maximum switching current 100mA q Maximum Ron 70 Ohm q Maximum switching frequency loadRL 240 VOUT 24Vdc 32 Hz q Insulation from other second...

Страница 8: ...me plate q Electric connections must be done as described at page 8 10 q Ground connections must be done Verify periodically q The integrity of the power supply cables wiring and other electrical part...

Страница 9: ...9 252_EN_IS_1_3_00 doc batteries pack Batteries Power Supply q Batteries quantity N 1 standard N 1 optional SWITCH ON DEVICE Switch battery 1 ON OFF SWITCH 1 2 INSTRUMENT START UP Switch battery 2...

Страница 10: ...user you can proceed programming all the functions up to L2 security level entering its code whenever you enter the Main menu ATTENTION take note very carefully of the customised code you have chosen...

Страница 11: ...ected decimals Beyond this value the totalise are reset Alarm ON Flow rate visualization Direct Reverse totalizer Flow rate and direct reverse totalizer visualization Speed of fluid Alternate pages Al...

Страница 12: ...CONVERTER LED INTERPRETATION Always switch off not power supply or trouble circuit Always switch on firmware not valid Continuos quickly flash normal operation in continuoe sampling mode Slowly flash...

Страница 13: ...G 1 SECOND It moves the cursor rightward on the input field It goes to the following subject of the menu It change the display of the process data LONG PRESSING 1 SECOND It moves the cursor leftward o...

Страница 14: ...2 Low flow zero threshold 0 25 of full scale value 3 4 Automatic change of scale 3 5 Energy saving 3 3 Enable every hour an internal cycle of calibration The measure it s stopped for 8 15sec 4 1 Maxim...

Страница 15: ...alizer difference between direct and reverse 8 9 Visualizes the values of the partial totalise in the unit of selected currency 8 10 Choice of the numbers of decimals for the visualization currency va...

Страница 16: ...r auto test 10 3 Flow rate simulation enabling 11 1 Level 2 access code enter 11 4 Load user data saved 11 5 Save user data 11 6 Visualisation of the total operation hours of the converter functionnot...

Страница 17: ...he principal functions q Through the Main menu where is possible to access to all function with access code 2 Below are brought some examples relating to the change of the value in the function 2 1 Fs...

Страница 18: ...m 4dm s to 5dm s from Main Menu quick start menu enable Enter in the Quick start menu Access to the Main Menu Access to the Scale menu Access to the function Fs1 Change the value Confirm the new value...

Страница 19: ...ou are sure the a m conditions are fulfilled press for more than one second the key will start one counter for 60 s after that check if the zero is correct otherwise repeat the operation again Press t...

Страница 20: ...2 8 2 9 Minimum maximum value for input 4 20mA I IS bar XXX XX Setting the minimum maximum value for external device with 4 20mA output There are four fields to fill in to set this parameter from lef...

Страница 21: ...to on off output 1 2 OUT 1 2 XXXXXX Choice of the function corresponding to digital Output 1 2 The functions are listed in the table below MENU 8 DISPLAY POS 8 2 8 3 8 4 8 5 Reset totalizer T P RESET...

Страница 22: ...following question EXECUTE press the key to proceed Press any other key to delete the operation POS 10 2 Autotest function enable SELF TEST Meter auto test function This function stops the normal fun...

Страница 23: ...ry data Any previous programming is cancelled getting back to the manufacturer s standard values POS 11 4 User pre settings loading LOAD USER PRES This function recalls the values saved from the user...

Страница 24: ...Check the status of the cables connecting the sensor to the converter CURR LOOP OPEN The 0 4 20mA output on board or the optional one are not correctly closed on a valid load Verify the load is applie...

Страница 25: ...IN ACCORDANCE WITH LOCAL REGULATIONS BATTERY SUBSTITUTION verify the absence of power supply set in stand by the instrument menu 10 extracts the battery group unscrewing the fixing screw suitable in...

Страница 26: ...26 252_EN_IS_1_3_00 doc...

Страница 27: ...27 252_EN_IS_1_3_00 doc...

Страница 28: ...252_EN_IS_1_3_00 doc The last three character of file name identify the sw version which the manual is refer the sw version is visualized during switch on of converter...

Отзывы: