The converter can be used in two different modes:
q
with continuous sampling
q
with sampling to preset unit of time.
q
CONTINUOUS SAMPLING
(pic.1)
(ENERGY SAVING OFF)
In this mode the converter make the measure in accordance with the classical diagram of the
flow meter; the consumption of the system, with any diameter of the sensor is 0,05 W ;
the
life of battery is about 1 month (6 with 6 battery)
The accuracy of the system is definite in those conditions.
q
SAMPLING TO PRESET UNIT TIME
( pic. 2 )
( ENERGY SAVING ON )
This mode works sampling the range to intervals of preset time (see MEASURE menu, func.
3.5); it allows a great saving of energy
In this conditions the consumptions are:
Sampling
time (s)
Battery life - n° 1
battery (years)
Battery life – “ N “
batteries (years)
1
0,7
0,7 * N
2
1.3
1.3 * N
5
2.1
2.1 * N
10
2.7
2.7 * N
15
3.0
3.0 * N
>= 30
5
5.0 * N
Whatever the results ,
the maximum time is
limited to 10 years
FLOW RATE
SAMPLE
FREQUENCY
SAMPLE
TIME
t
t
FLOW RATE
SAMPLE
FREQUENCY
PIC 1
PIC 2
ATTENTION:
the consumptions on the table are without optional IF2 cable connected to the
converter. This device accelerate the consumption of the batteries even if the instrument is in standby
mode
. Recommends to disconnect the IF2 cable from the converter after every his uses or
switch off the converter by the dip-switches (see page 9 )
MEASURE/ CONSUMPTIONS