21 / 89
18 OUT 1(+)
19 OUT 2(+)
17 (COM. OUT 1-2)
mA
24 OUT 1(+)
25 OUT 2(+)
23 (COM. 4-20 OUT 1-2)
OUTPUTS WIRING
digital outputs
analog outputs
Opto-insulated output with floating collector and emitter
terminals freely connectable
Maximum switching voltage: 30V
Maximum switching current: 100mA @ 25°C
Maximum saturation voltage between collector and emitter
@100mA: 1.2V
Maximum switching frequency (load on the collector or
emitter, RL=470, VOUT=24V
): 1250Hz
Maximum reverse current bearable on the input during and
accidental polarity reversion (VEC): 100mA
Insulation from other secondary circuits 500 V
Opto-insulated output
Maximum load: 1000Ω
Maximum voltage without load: 27V
Refresh frequency is the same of the sample frequency
of the connected sensor
Protected against persistent over voltages to maximum
30V