background image

11 

 THIS INSTRUMENT MUST NOT BE USED ON UNINSULATED CONDUCTORS AT A VOLTAGE GREATER THAN  

       600V ac/dc. 

 
3-2 AC Current Measurement 

 
1. Set the AC/DC switch to the AC

 position. 

2. Set the function switch to the desired range, if unsure select 600A range. 
3. Press the trigger to open the current transformer jaws and clamp around one conductor only, making sure that the jaw is  
    firmly closed around the conductor, then take the reading from the digital display. 

 

3-3 DC Current Measurement 

 
1. Set the AC/DC switch to the DC

 position. 

2. Set the function switch to the desired range, if unsure select the 600A range. 
3. Adjust the "DCA Zero Adj. Knob" until the display shows "0". 
4. For  DC measurement, the reading is a positive value when the current flows from the lower side to the upper side of the  
     instrument as Fig. 2. 
5. Press the trigger to open the transformer jaws and clamp one conductor only, making sure that the jaw is firmly closed  
    around the conductor, then take the reading from the digital display. 

Содержание ICM36R

Страница 1: ...INSTRUCTION MANUAL ICM36R CLAMP MULTIMETER ICM 36R EN FR IT DE...

Страница 2: ...1 ISO TECH ICM36R CLAMP MULTIMETER INSTRUCTION MANUAL...

Страница 3: ...2...

Страница 4: ...following items 1 Current meter 2 Carrying case 3 Instruction Manual 1 2 Meter Safety Terms as Marked on Equipment Attention Refer to Manual Double Insulation Protection Class II DANGER Risk if electr...

Страница 5: ...from slipping off the hand while in use 3 Function Switch Slide switch is used to select the desired range 4 DC Zero Adj Knob Zero adjusts on DC current ranges 5 AC DC Slide Switch Selects a c or d c...

Страница 6: ...5 Figure 1...

Страница 7: ...reading of 1999 Polarity Automatic polarity indicated Overrange Indication 1 or 1 indicated Measuring Rate 2 times per second nominal Position Error 1 of reading Type of Sensing Hall effect sensing f...

Страница 8: ...ient 0 C 50 C 75 R H Storage Temperature 20 C 60 C 80 R H Temperature Coefficient 0 2 x Specified accuracy C 18 C or 28 C Power Requirement Alkaline 9V battery Battery Life 40 hours typical Alkaline M...

Страница 9: ...ications Accuracy is reading number of digits at 23 C 5 C less than 75 R H 1 DC A Range Resolution Accuracy 200A 0 1A 1 9 reading 5digits for 0 20A 1 9 reading 3digits for 20A 200A 600A 1A 1 9 reading...

Страница 10: ...20A 1 9 reading 5digits for 20A 200A 600A 1A 1 9 reading 5digits for 200A 600A Frequency Response 40Hz 400Hz Data Hold Hold display reading for all functions and ranges 3 Auto Power Off The meter will...

Страница 11: ...1 Make sure that the battery is properly connected 2 The instrument should only be operated between 0 C 50 C and at less than 75 R H 3 Do not use or store this instrument in a high temperature or high...

Страница 12: ...onductor then take the reading from the digital display 3 3 DC Current Measurement 1 Set the AC DC switch to the DC position 2 Set the function switch to the desired range if unsure select the 600A ra...

Страница 13: ...12 Figure 2...

Страница 14: ...ly wipe the case with a damp cloth and detergent do not use abrasives or solvents Battery Replacement This instrument is powered by a single 9V alkaline battery Use the following procedure to replace...

Страница 15: ...nited Kingdom Japan West Tower 12th Floor Yokohama Business Park 134 Godocho Hodogaya Yokohama Kanagawa 240 0005 Japan USA 7410 Pebble Drive Fort Worth Texas 76118 6961 Canada 1701 Woodward Drive Ste...

Отзывы: