5.2.  Sensitization test

 

Before carrying out your first treatment you must carry out a 

sensitization test. Follow the step by step guide below;
1. 

Place the décoLITE over your inner arm.

2. 

Switch ON the controller and allow the décoLITE to run  

for its full 10-minute treatment program.

3. 

At any time if your skin feels HOT or uncomfortable  

STOP using the device immediately.

4. 

After the treatment wait at least 6 hours.  If your skin on  

your inner arm is red or itchy then the décoLITE is unsuitable 

for you. Return the device to the retailer within (14) days  

of purchase.

6.0.  Using décoLITE Device

6.1.  Treatment Recommendations:

 

• 

For best results treat for 3-5 days a week for a minimum  

of 4 weeks.

• 

Maintenance treatment should be carried out as required, 

according to lifestyle and skin condition.

6.2.  Wearing the décoLITE LED system

• 

décoLITE comes with an adjustable Velcro strap that allows 

you to comfortably wear the device in two positions, around 

the neck and décolletage or décolletage only 

• 

To fit the mask, thread the shorter straps through the inlet 

holes on either side of the device for neck and décolletage.  

Alternatively, to treat a larger area of the décolletage put 

the longer strap through the holes at the top of the device.

• 

Loop the straps around your neck and secure firmly.

• 

Adjust if necessary.

6.3.  Carrying out a treatment

1. 

Cleanse and exfoliate neck and décolletage to remove any 

traces of skincare

3. 

Place décoLITE on the neck and décolletage, secure with 

the Velcro strap and switch on the controller.  The device 

switches off after a ten-minute treatment time. You may stop 

the treatment at any time by pressing the ON/OFF button 

on the controller.

4. 

Remove the mask, apply your normal skincare and 

sunscreen.

5. 

Clean the mask using a water-based wipe or a damp clean 

cloth.  Ensure that the device is NOT being charged when 

cleaning.

6. 

Repeat 3 – 5 times a week for four weeks.

7.0.  Care and Maintenance

• 

Store décoLITE LED device at room temperature in dry 

conditions. Use the carry bag provided.

• 

Avoid storing the décoLITE LED system in dusty conditions, in 

direct sunlight or in damp and humid conditions.

• 

The décoLITE LED system contains no serviceable parts.

• 

Before each use, with the device switched off, clean the face  

mask following the cleaning instructions below.

7.1.  Cleaning the décoLITE LED system

  Do NOT

 use any other cleansing products as these may  

damage the plastics.
• 

Clean the transparent side (side in contact with the skin) of 

the mask and the rear of the mask with a water-based wipe 

or damp clean cloth. Place the mask on a clean dry surface 

to air dry. Visually inspect the surfaces to ensure they are 

clean. Stubborn stains can be removed by wiping gently 

with 70% alcohol.

7.2. Disposal

  DO NOT

 dispose of any part of the device at a domestic  

waste facility. 

 

The décoLITE controller contains a Lithium ion polymer battery. 

For protection of the environment, please package and return 

this unit to the manufacturer or their representative at the end  

of its working life or safely dispose of the device following your 

local guidelines. 

7.3. Troubleshooting

Problem

Possible Cause

Action

No red or 
infra-red light 
from décoLITE

Not charged

Plug in the controller to the 
charger making sure that the 
connectors are attached and 
leave charging until the three 
indicator LEDs on the controller 
remain lit and do NOT flash

No red or 
infra-red light 
from décoLITE

Device 
connected to 
the mains

Unplug micro USB connector 
from controller

LEDs do NOT 
turn OFF after 
10 minutes

Device is faulty

Return to your supplier

LEDs flicker 
during 
treatment

décoLITE USB 
is not fully 
inserted into the 
controller

Ensure the controller and 
décoLITE are connected 
correctly

Содержание Natural facelift decoLITE

Страница 1: ...USER GUIDE ...

Страница 2: ...E LED system or this User Guide décoLITE and the Natural facelift Company are trademarks of iSMART Developments Ltd Registered office Fulford House Newbold Terrace LEAMINGTON SPA CV32 4EA United Kingdom hello natural facelift com www natural facelift com The information in this document is proprietary to iSMART Development Ltd and supplied on the understanding that it shall be kept confidential to...

Страница 3: ...r smoother and with a visible improvement in skin tone and texture lines and age spots 1 1 Intended use The décoLITE LED system emits light in the red and near infra red NIR region of the light spectrum and can be used to help reduce the signs of aging It is also indicated for the alleviation of pain increased blood flow and temporary relaxation of muscle can provide pain relief 2 0 Device Descrip...

Страница 4: ...se décoLITE LED system to treat any other conditions apart from those listed in the indications for use The décoLITE LED system has not been tested for any other conditions than those listed and the risk is unknown DO NOT use décoLITE LED system if you are pregnant planning on becoming pregnant or breastfeeding The décoLITE LED system has NOT been tested on pregnant or breast feeding women and the...

Страница 5: ...ld present an electrical shock hazard to the user Please read the accompanying user manual thoroughly and follow directions stated in the manual to assure maximum safety during operation Precautions The décoLITE strap contains a mixture of Polyester nylon and lycra If you are allergic to any of these materials you may not be able to use the straps provided The décoLITE system CANNOT be used in cha...

Страница 6: ...nd sunscreen 5 Clean the mask using a water based wipe or a damp clean cloth Ensure that the device is NOT being charged when cleaning 6 Repeat 3 5 times a week for four weeks 7 0 Care and Maintenance Store décoLITE LED device at room temperature in dry conditions Use the carry bag provided Avoid storing the décoLITE LED system in dusty conditions in direct sunlight or in damp and humid conditions...

Страница 7: ...rds for non laser light source medical equipment for home use EN 60601 1 11 2015 EN 60601 2 57 2011 and EN 60601 1 2 2015 and IEC 62471 2006 Photobiological Safety of Lamps and Lamp Systems EN 55014 1 2006 A1 2009 A2 2011 Electro Magnetic Compatibility requirements for household appliances electric tools and similar apparatus Part 1 emissions EN 55014 2 2015 Electromagnetic compatibility requireme...

Страница 8: ...ght to replace a defective product with the most compatible product currently available This warranty does not cover damage caused by the following Accident misuse abuse or alteration Servicing by unauthorized persons Connection to incorrect power sources Use with unauthorized accessories iSMART Developments Ltd or its authorised distributors shall not be responsible for any incidental special or ...

Страница 9: ...ssential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC EN 60601 2 57 2011 Medical electrical equipment Particular requirements for the basic safety and essential performance of non laser light source equipment intended for therapeutic diagnostic monitoring and cosmetic aesthetic use EN 62471 ...

Отзывы: