![IOtech 652u Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/iotech/652u/652u_user-manual_2088021008.webp)
1-2 What are the IOtech 652u and 655u?
917891
IOtech 652u and 655u User’s Manual
Features of the Dynamic Signal Inputs
o
a current source for transducer biasing (IEPE)
o
detection of a transducer fault
o
AC coupling: 0.1 Hz; or DC coupling
o
±40 V range
o
anti-aliasing filters: 3-pole low pass filter in hardware;
software FIR filters are automatically set for each analysis rate
o
pseudo-differential inputs
o
support for TEDS (Transducer Electronic Data Sheet) in eZ-Analyst
o
any analog input channel can serve as a tach input
o
tight channel-to-channel phase matching
USB2.0 ready
:
Easy Connection to
USB2.0-ready
Notebooks, Desktop PCs, or USB2.0 Hubs
.
Note that the USB2.0 port allows a continuous stream of data to be collected and stored in the host PC.
Analog Input Channels
:
10 BNC connectors
Temperature Measurement Channels (655u only)
:
5 channels, thermocouple or RTD. Wiring
connects to 655u rear panel via a removable connector block.
8 Digital I/O Channels
: DB9 connector for connection of Digital I/O signal lines.
Note
: eZ-NDT or eZ-TOMAS software must be used to make use of the Digital I/O.
Analog Channel Triggering
Pre- and Post-Trigger Readings
Reference Notes
:
For information regarding 655u connections for temperature measurement, refer to chapter 3.